Product Service 1-800-522-7658
28
QuickChart – Logic+ Type B2 Metric Units
Model
30348
30368
Type
B2
B2
Cabinet Size (in m)
1.22
6
Sash Opening (mm)
203
8
Starting Serial #
1303_
1303_
Downflow Data
Nominal Avg. Downflow (m/s)
.279+/.03
55+/-5
grid # of points (rows x columns)
24 (3x8)
36(3x12)
Grid distance from back & sides
152
6.0
Distance between rows (mm)
146
5.75
Distance between columns (mm)
132
5.50
Inflow Data
Nominal Average Inflow (m/s)
.533+/-.03
105+/-5
Sash Open Area (Sq. m)
.250
.374
Nominal Avg. Exhaust Vol. (m3/s)
.134
.200
Avg. Inflow Vol. Range (m3/s)
.127-.140
.190-.209
Secondary Inflow Data
3
Sash Opening Template
1
4+
6+
Sensor distance (mm)
2
83
4
# of Test points
8
12
Test point location
*
*
Nominal Avg. Inflow Vel. (m/s)
1.44
1.17
Avg. Inflow Vel. Range (m/s)
1.37-1.50
1.12-1.23
Correction Factor (CF)
.0929
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.127-.140
.190-.209
Sash Open Area (Sq. m)
.250
.374
Inflow Velocity Range
(AIV / Sash open area)
.533+/-.03
.533+/-.03
Supply HEPA Leak Test Data
Air Displacement (m3/s)
.210
.471
Laskin Nozzles needed
1
1
Theoretical aerosol conc. (ug/l)
3
30
20
Actual aerosol conc. (ug/l)
4
15
12
Exhaust HEPA Leak Test Data
Air Displacement (m3/s)
.314
.471
Laskin Nozzles needed
2
2
Theoretical aerosol conc. (ug/l)
3
41
27
Actual aerosol conc. (ug/l)
4
25
19
Exhaust Data
DIM Exhaust Volume (m3/s)
5
.314
.471
Traverse Exhaust Volume (m3/s)
6
.382
.575
Differential Pressure (Pa)
7
374
448
Exhaust (Inflow) Alarm Data
Alarm Setpoint (m3/s Inflow)
.071 +/- .03
.105 +/-.03
Supply HEPA Data
Width x Depth x Height (in.)
8
48x18x3.06
72x18x3.06
Performance (CFM)
445
665
Performance (Pressure in. H
2
O)
.5+/-.1
<.42
Labconco P/N
3838401
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
8
26x18x8.08
48x18x8.08
Performance (CFM)
720
1100
Performance (Pressure in. H
2
O)
.48+/-.07
.37+/-.07
Labconco P/N
3438501
3438503
Motor/Blower Data
Labconco P/N
6
3832209
3832210
Motor HP
½
½
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721903
Fluor. Ballast Labconco Part #
3838100
3838100