Calibration Procedure-
NOTE: YOU MUST FOLLOW THESE INSTRUCTIONS EXACTLY, OR THE SENSOR
WILL NOT OPERATE PROPERLY!
Note: You will need to access the certifiers menu to calibrate the airflow
sensor.
Note: To properly calibrate the air flow sensor, a qualified certifier must
verify the inflow and downflow values.
With the cabinet operating:
8.
Press the
Menu
button.
9.
Press the
▼
button until the
Services
option is highlighted
(it will turn yellow).
10.
Press
OK/Mute
to enter the Services menu screen.
11.
Press the
▼
button until the
Calibration
option is
highlighted (it will turn white).
12.
Press
OK/Mute
to start the calibration procedure.
13.
You will see a
warning screen
alerting you that you are
about to alter the BSC’s settings.
14.
Press
OK/Mute
to input the password.
NOTE: IF THE UNIT IS CONNECTED TO AN EXHAUST SYSTEM, YOU MUST
STOP ALL AIR FLOW OVER THE SENSOR DURING THE INITIAL ZEROING
OF THE SENSOR - DO THIS BY SHUTTING OFF THE EXHAUST SYSTEM, OR
BLOCKING THE SENSOR INTAKE WITH TAPE.
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Configuration
Calibration
Service
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Calibration
OK to proceed, MENU to return
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape