Moving and Lifting the Cabinet
Move the cabinet, attached to its pallet, by using a floor jack, or a furniture dolly underneath the
unit. DO NOT move the cabinet by tilting it onto a hand truck.
When lifting the cabinet DO NOT lift the cabinet in the middle front area of the hull. Lifting here
may bend or distort the bottom of the cabinet, causing damage to the unit.
Installing the Biosafety Cabinet on an
Existing Work Surface
Note: The biosafety cabinet is very top heavy. Use caution when lifting or moving it.
When installing the biosafety cabinet onto an existing work surface or benchtop, ensure
that the structure can safely support the combined weight of the cabinet and any related
equipment. The work surface should be at least as wide as the cabinet and 31 inches (787
mm) deep to properly support the unit.
A hole or notch may be cut in the supporting surface in the right front corner to
accommodate the optional drain valve.
Installing the Cabinet on a Labconco Base
Stand
Labconco offers accessory Base Stands in a variety of configurations to suit your
particular needs. Stands can be ordered with adjustable telescoping legs or with a
manually or electrically adjustable hydraulic lift.
Telescoping Base Stands
These stands are included with some Logic+ models, or are available separately. The
base stands for each width cabinet are listed in Table 3-1 below. An optional caster
wheel kit is available (part # 3730500).