7
TECHNICAL DATA
78
OPTISYS SLM 2100
www.krohne.com
01/2017 - 4002737304 - MA OPTISYS SLM 2100 R04 en
7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Measuring principle
Level measurement via immersion of optical sensor with straight
light transmission of NIR-light for suspended solids measurement.
Application range
Level measurement of sludge blanket, fluff zone and zone tracking in
clarifiers, sedimentation basins as well as in thickeners of water and
wastewater treatment plants.
Design
Measuring range
0.1...30 g/l (depending on the sludge type)
Units
Concentration in mg/l or g/l
Height / depth
m, cm, feet, inch
Lowering speed sensor
Maximal 7.75 cm / 3.05" per second at 50 Hz
Full cycle time for 10 m / 32.8 ft: 3 min
Internal heating
"Off" above 8
°
C / 46.4
°
F, full power below 4
°
C / 39.2
°
F.
Water connection of the cleaning
unit (optional)
3/4" connector with metric thread (max. pressure 6 bar / 87 psi)
Display and user interface
Graphic display
LCD display, backlit 128 x 64 pixels, temperature below -25
°
C / -13
°
F
may affect the readability of the display
Operating elements
4 push buttons for operator control of the signal converter without
opening the housing
Operating menu
The operation menu consists of the measuring mode and the menu
mode
Measuring mode
2 pages - measuring page with measuring results and status page
with status messages
Menu mode
Variety of main and submenus that allows customising the device
Operating and display languages
English, German, French and Spanish
Summary of Contents for OPTISYS SLM 2100
Page 85: ...NOTES 8 85 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Page 86: ...8 NOTES 86 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Page 87: ...NOTES 8 87 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...