5.8.2 Error category "Application error" ( "F", not bold)
5.8.3 Error category "Information" (I)
5.8.4 Error category "Out of specification" (S)
INFORMATION!
Regard that also in case of an error message of the category "F" the measured values are not
valid!
Messages on the display Description
Actions
F Door open
The door of the cable drum compartment is
open. Motor stops and measurement cycle
is interrupted.
Close the door; if closed check safety
switch; Error will be automatically reset, if
problem was solved.
F Sensor supply voltage
Voltage supply out of range.
Check if pick up axle board is correctly
positioned and the optical path is clean.
Error needs to be reset manually and
measurement can be started again. The
distance for correct transmission should
not exceed 2 mm / 0.079 inch.
Messages on the display Description
Actions
I Search reference point
Sensor returns to home position to set zero
point of level measurement.
No action necessary!
I Motor up
Motor reels up.
I Motor down
Motor reels down.
I Heater active
Internal heater is switched on.
I Fan active
Fan is switched on.
I Concentration not found
The sensor did not detect the concentration
set by the user during this measuring cycle.
I Cable cleaning active
Cleaning unit is switched on.
I Sensor cleaning active
Cleaning unit is switched on.
I Rake guard
Rake guard switch is active.
I Pump lead time
Pump lead time for external pump is active.
I Relay 1 active
Indicating that relay 1 is active.
I Relay 2 active
Indicating that relay 2 is active.
INFORMATION!
Regard that in case of an error message of the category "S" the measured values are only
conditionally usable!
Messages on the display Description
Actions
S Sensor comm. warning
The communication to the sensor was
interrupted.
Warning, that communication to sensor was
interrupted for a short time. No action
necessary. Error will be automatically reset,
if error does not occur again.
S Out of spec.
concentration
Current output out of range (valid in Profile /
tracking mode for current channel B,
proportional to the measured
concentration).
Check the limits of the 4
…
20 mA current
outputs. If the maximum concentration is
exceeded please enter a larger value.