C2, maintenance
C3, I/O
C1.6
tracking time
Definition of the time period the sensor is remaining in parking position,
when elapsed the sludge level will be tracked again (this menu is displayed
only if "zonetracking" is selected in C1.1).
Range: 20
…
999 seconds; Default: 30 seconds
C1.7
tracking position
Definition of the relative parking position of the sensor above the sludge
level (this menu is displayed only if "zonetracking" is selected in C1.1).
Range: 0.05...0.3 m; Default: 0.15 m
C1.8
start signal
This menu shows the start signal for the measurement and can be set to
"Timer" (via build in timer) or "Trigger input" (via external switch on control
input 1 and/or 2). This menu is displayed only if "level" or "profile" is
selected in C1.1
C1.9
timer interval
This menu shows the timer interval for starting the measurement and can
be set in minutes (this menu is displayed only if "timer" is selected in C1.8.
and "level" or profile in C1.1). Range: 1...999 min; Default: 10 min
C1.10
units
This menu groups several functions to define the units of the measuring
results. These units are used for all values of the related unit group on the
display (measuring screen and parameters).
C1.10.1
height / depth
This function allows to define the unit for level measurement.
Options: m, feet, inch, cm
C1.10.2
temperature
This function allows to define the temperature unit. Options:
°
C,
°
F, K
C1.10.3
concentration
This function allows to define the concentration unit. Options: g/l, mg/l
Menu level Designation / function
Settings / descriptions
This menu groups several other menus in which the maintenance functions can be set.
C2.1
calibration
In this menu the concentration measurement can be calibrated. For further
information refer to
Calibration
on page 65.
C2.2
ref. position calib.
In this function the reference point calibration of the height / depth
measurement can be set. Depending on the application this might be
necessary due to dirt accumulation on the cable.
Options: off or time interval
Default: off
C2.2.1
activation
This function allows to activate the ref. position calibration.
Options: off or timer interval
Default off
C2.2.2
timer interval
Timer interval of reference point calibration. This menu is only displayed if
"time interval" is selected in C 2.2.
Range: 2...999 hours
Default: 1 hour
Menu level Designation / function
Settings / descriptions
This menu groups several other menus in which the I/O functions can be set.
C3.1
current out A
Settings of the corresponding current output. In height/depth mode, the
current output is calculated automatically.
C3.1.4
height / depth 4 mA
Indicates the height/depth for 4 mA, calculated based on entered basin
geometries - read only
C3.1.5
height / depth 20 mA
Indicates the height/depth for 20 mA, calculated based on entered basin
geometries - read only.
Menu level Designation / function
Settings / descriptions