5
OPERATION
52
OPTISYS SLM 2100
www.krohne.com
01/2017 - 4002737304 - MA OPTISYS SLM 2100 R04 en
5.6 Function tables
5.6.1 Menu A, quick setup
A1, language
A2, set clock
A3, reset errors
A4, measuring function
5.6.2 Menu B, test
The procedure to start the simulation process is the same for all functions:
•
Choose the function using
↑
or
↓
and press >.
i
You see the two options set value
set value
set value
set value (opens the editor to enter the simulation value) and break
break
break
break
(exit the menu without simulation).
•
Choose the desired option using
↑
or
↓
and press >.
i
If you choose break
break
break
break you will exit the menu without simulation. If you chose set value
set value
set value
set value, the
device asks start simulation
start simulation
start simulation
start simulation and offers the options no
no
no
no (exit the menu without simulation) or
yes
yes
yes
yes (starts the simulation finally).
•
Choose the desired option using
↑
or
↓
and press >.
i
If you choose yes
yes
yes
yes, the simulation starts. If you choose no
no
no
no, you will exit the menu without
simulation.
B1, simulation
B2, actual values
Language selection (depends on the region for which the device has been ordered). Available languages: Deutsch,
English, Français, Español
Manual setting of date and time.
This functions allows to reset all errors that are not reset automatically (e.g. motor blocked). You can answer the
question "reset?" with the following options: no (exit without reset) / yes (reset and exit the function).
This menu shows the different measurement functions of the sludge level meter. Choose between: "level", "profile",
"zonetracking" and "manual operation". For further information refer to
Measuring function
on page 61.
Menu level Designation / function
Settings / descriptions
This menu groups several functions to simulate the input and output values. The procedure to start the simulation
process is the same for all functions.
B1.1
B1.1
B1.1
B1.1
current out A
Allows simulation of the corresponding value. It might be useful during
installation.
B1.2
B1.2
B1.2
B1.2
current out B
B1.3
B1.3
B1.3
B1.3
relay 1
B1.4
B1.4
B1.4
B1.4
relay 2
B1.5
B1.5
B1.5
B1.5
cleaning
Menu level Designation / function
Settings / descriptions
This menu groups several functions which allow to display the corresponding actual reading.
B2.1
B2.1
B2.1
B2.1
operating hours
This menu shows the operating time of the devices in hours.