276
The system displays the Mechanical and Thermal Indices during real-time imaging, in all imaging modes .
To check MI/TI values, the image on the study screen needs to be in a live state .
• MI/TI values are hidden during frozen state . Press the
FREEZE
button to cancel the frozen state .
• MI/TI values are dependent on mode and [Tx Level] changes .
Tx Level
Mechanical and Thermal Indices display area
Indices are displayed in the abbreviated form as shown below:
[MI] : Mechanical Index
[TIS] : Soft Tissue Thermal Index
[TIB] : Bone Thermal Index (or [TIC] : Cranial Thermal Index)
9 .4 Display Layout of MI/TI Values
Summary of Contents for SONIMAGE HS2
Page 1: ...EN 01...
Page 2: ......
Page 6: ...6...
Page 7: ...7 Introduction...
Page 41: ...41 For all product inquiries contact your service representative Support...
Page 42: ...42...
Page 43: ...43 Chapter 1 Product Overview...
Page 68: ...68...
Page 69: ...69 Chapter 2 Before Examination...
Page 84: ...84...
Page 85: ...85 Chapter 3 Starting Examination...
Page 122: ...122...
Page 123: ...123 Chapter 4 Diagnosis Mode...
Page 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Page 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Page 232: ...232...
Page 233: ...233 Chapter 7 Maintenance Inspection...
Page 243: ...243 Chapter 8 Transducer and Biopsy...
Page 270: ...270...
Page 271: ...271 Chapter 9 MI TI...
Page 283: ...283 Chapter 10 Acoustic Output...
Page 286: ...286...
Page 287: ...287 Chapter 11 Others...
Page 305: ...305...
Page 306: ...306...
Page 307: ......