175
Chapter 4
4 .8 Panoramic View
The panoramic view is a function that displays a panoramic image by combining the B-mode image with the motion
of the transducer .
The transducers that can use it are the L18-4, L14-4, and L11-3 .
You must have the optional "Panoramic View License" to use this function . Contact your Konica Minolta service rep-
resentative .
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• Panoramic images are generated based on continuous multiple images captured by scanning, whereupon the posi-
tion of each image is estimated and combined according to those results . Because of this, distortion may occur, due
to the scanning, in the panoramic image that is captured . Use panoramic images as reference images .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
4 .8 .1 Capturing Panoramic Images
1
Press the [Menu] button > [Panoramic View] button on the [Other] tab .
• The panorama setup screen appears, and then the panoramic ROI appears .
• The depth is in a range from 3 to 7 cm .
• The screen switches to the B-mode image .
Panoramic ROI
Summary of Contents for SONIMAGE HS2
Page 1: ...EN 01...
Page 2: ......
Page 6: ...6...
Page 7: ...7 Introduction...
Page 41: ...41 For all product inquiries contact your service representative Support...
Page 42: ...42...
Page 43: ...43 Chapter 1 Product Overview...
Page 68: ...68...
Page 69: ...69 Chapter 2 Before Examination...
Page 84: ...84...
Page 85: ...85 Chapter 3 Starting Examination...
Page 122: ...122...
Page 123: ...123 Chapter 4 Diagnosis Mode...
Page 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Page 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Page 232: ...232...
Page 233: ...233 Chapter 7 Maintenance Inspection...
Page 243: ...243 Chapter 8 Transducer and Biopsy...
Page 270: ...270...
Page 271: ...271 Chapter 9 MI TI...
Page 283: ...283 Chapter 10 Acoustic Output...
Page 286: ...286...
Page 287: ...287 Chapter 11 Others...
Page 305: ...305...
Page 306: ...306...
Page 307: ......