4 .4 Basic Scanning Mode
154
CAUTION
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• The [Stable Section] function displays the estimated interval in which stable pressurization takes place . It does not
display the diagnostic image .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• The following B-mode parameters change automatically when you enter Elastography-mode . The original values are
automatically restored when you exit Elastography-mode .
– [B Density] : If higher than [H], select [H]
–
[Trapezoid] : [Off]
– [Focus Pattern] : 1
• The following settings change automatically when you enter Elastography-mode . The original state is not restored
when you exit Elastography-mode .
–
[Biopsy], [SNV] : [Off]
–
[ECG] : [Off]
•
When using Elastography-mode, M-mode/PW Doppler-mode/PW-TDI/Color flow/Power Doppler/SCF-mode cannot
be activated .
• When using Dual-mode, if one of the planes is in Elastography-mode, Biopsy, and ECG cannot be activated .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•
Except when [Off] is selected for [Equalization] uneven coloring can be alleviated.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Pressurization state navigation
Pressurization state navigation is a function for assisting the elastic image display operation .
To display the elastic image continuously, repeat following action: lightly pressurize the target site with the transducer
and then release the pressure .
Pressurization state navigation assists the operation by displaying the pressurization state in three ways (Navigation
types) .
z
Navigation type 1
In Navigation type 1: The strain time variation is displayed by line chart .
If the line chart stays between the upper and lower blue lines, the elastic image can easily be displayed by re-
peating the pressurization and relaxation .
z
Navigation type 2
Minimum score display example
Maximum score display example
Navigation type 2 displays the pressurization and relaxation repetitions as scores .
The higher the score the more easily the elastic image is displayed .
Summary of Contents for SONIMAGE HS2
Page 1: ...EN 01...
Page 2: ......
Page 6: ...6...
Page 7: ...7 Introduction...
Page 41: ...41 For all product inquiries contact your service representative Support...
Page 42: ...42...
Page 43: ...43 Chapter 1 Product Overview...
Page 68: ...68...
Page 69: ...69 Chapter 2 Before Examination...
Page 84: ...84...
Page 85: ...85 Chapter 3 Starting Examination...
Page 122: ...122...
Page 123: ...123 Chapter 4 Diagnosis Mode...
Page 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Page 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Page 232: ...232...
Page 233: ...233 Chapter 7 Maintenance Inspection...
Page 243: ...243 Chapter 8 Transducer and Biopsy...
Page 270: ...270...
Page 271: ...271 Chapter 9 MI TI...
Page 283: ...283 Chapter 10 Acoustic Output...
Page 286: ...286...
Page 287: ...287 Chapter 11 Others...
Page 305: ...305...
Page 306: ...306...
Page 307: ......