1 .2 Safety Precautions
CAUTION
(7) EMC Statement
Guidance and manufactureís declaration - electromagnetic emissions
The unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit should assure that it is used in such an environment .
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The unit uses RF energy only for its internal function .
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment .
RF emissions
CISPR 11
Class B
The unit is suitable for use in all establishments other than
domestic and those directly connected to the public lowvoltage
power supply network that supplies buildings for domestic purposes .
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
(8) EMC Statement
Guidance and manufacture’s declaration - electromagnetic immunity
The unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit should assure that it is used in such an environment .
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 6 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humid-
ity should be at least 30 % . Mains
power quality should be that of a typi-
cal commercial or hospital environ-
ment .
+/- 8 kV air
+/- 8 kV air
Electrical fast transient/
burst
IEC 61000-4-4
+/- 2 kV for power sup
-
ply lines
+/- 2 kV for power sup
-
ply lines
+/- 1 kV for input/output
lines
+/- 1 kV for input/output
lines
Surge
IEC 61000-4-5
+/- 1 kV differential mode +/- 1 kV differential mode
Mains power quality should be that of
a typical commercial or hospital envi-
ronment .
+/- 2 kV common mode +/- 2 kV common mode
Voltage dips, short in-
terruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0 .5 cycle
<5% U
T
(>95% dip in U
T
)
for 0 .5 cycle
Mains power quality should be that
of a typical commercial or hospital
environment . If the user of the unit
requires continued operation during
power mains interruptions, it is recom-
mended that the unit be powered from
an uninterruptedly power supply or a
battery .
40% U
T
(60% dip in U
T
)
for 5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(<95% dip in U
T
)
for 5 sec
<5% U
T
(<95% dip in U
T
)
for 5 sec
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment .
[NOTE] U
T
is the a .c . mains voltage prior to application of the test level .
Summary of Contents for REGIUS SIGMA2
Page 1: ...EN 05 0197 Operation Manual DIRECT DIGITIZER REGIUS SIGMA2...
Page 2: ......
Page 5: ...5 Introduction...
Page 10: ...10...
Page 21: ...21 Chapter 2 Product Outline An explanation of the device s features and functions...
Page 27: ...27 Chapter 3 Reader Operation An explanation on how to operate this device...
Page 35: ...35 Chapter 4 Cassette Plate Operation An explanation on cassette and plate handling...
Page 45: ...45 Chapter 5 Troubleshooting An explanation on how to deal with errors when they occur...
Page 62: ...62...
Page 70: ...70...
Page 71: ...71 Appendix A list of device specifications etc...
Page 74: ...74...
Page 75: ......
Page 76: ...A5ERBA01EN05 2015 08 31 JD...