Installing the PROCESSOR
981088 – September 1995
31
Installing the KNOBS
[83] Install the KNOBS for the DRAIN VALVES by doing the steps below:
(a) Check that all 3 DRAIN VALVES are closed.
(b) Install the 3 KNOBS.
Check that you align the SETSCREW of the KNOB with the flat section of the STEM.
Leave a 3 mm (
1
⁄
8
in.) gap between the FRONT ACCESS PANEL and the back of the KNOBS.
Secure the KNOBS using the 2 SETSCREWS provided for each.
Figure 25
Installing the KNOBS for the DRAIN VALVES
[84] Install the DRYER TEMPERATURE CONTROL KNOB.
(a) Align the SETSCREW of the KNOB with the flat section of the STEM.
(b) Press the DRYER TEMPERATURE CONTROL KNOB onto the STEM of the DRYER
TEMPERATURE CONTROL.
(c) Check that the DRYER TEMPERATURE CONTROL KNOB is not touching the FRONT ACCESS
PANEL.
[85] Advance to “Doing the Water Leak Test” procedure on Page 54.
Figure 26
Installing the DRYER TEMPERATURE CONTROL KNOB
H130_0149AA
PANEL
FRONT ACCESS
H130_0149ACA
KNOB (3)
SETSCREW
(2)
STEM (3)
+
-
Ready
Wait
Service
Dryer
Temperature
H130_0003AA
DRYER TEMPERATURE
CONTROL KNOB
H130_0003ACB