INSTALLATION INSTRUCTIONS
24
September 1995 – 981088
Routing the REPLENISHMENT HOSES
If the REPLENISHMENT TANKS are inside the darkroom, you need to route the REPLENISHMENT
HOSES through the darkroom wall. Do Steps 46 through 48. If you do not need to route the REPLENISHMENT
HOSES through the darkroom wall, advance to Step 49 on Page 25.
[46] Connect the 2 black, opaque
3
⁄
8
in. REPLENISHMENT HOSES provided in the THROUGH-THE-WALL KIT
to the clear REPLENISHMENT HOSES coming from the PROCESSOR.
Figure 19
Connecting the REPLENISHMENT HOSES
Note
When routing the HOSES through the MOUNTING STAND:
• check that you do not interchange the position of the HOSES between where they enter the MOUNTING
STAND and where they pass through the wall
• install the REPLENISHMENT HOSES with slack in the HOSES
• note that the developer REPLENISHMENT HOSE has a red WIRE TIE for easy identification
• note that the fixer REPLENISHMENT HOSE has a blue WIRE TIE for easy identification
• install the DRAIN HOSES without slack in the HOSES
[47] Route the following HOSES through the holes in WOOD FRAME, as needed:
• 3 DRAIN HOSES
• HOSE from the main water supply
• 2 REPLENISHMENT HOSES
[48] If you do not need to route the REPLENISHMENT HOSES through the darkroom wall, continue with the steps
below Otherwise, advance to Step 51 on Page 25.