Installing the PROCESSOR
981088 – September 1995
25
Note
When connecting the REPLENISHMENT HOSES to the REPLENISHMENT TANKS, install the STRAINER
ASSEMBLIES onto the REPLENISHMENT HOSES near the REPLENISHMENT TANKS.
[49] Use the parts from the THROUGH-THE-WALL KIT listed below to route the REPLENISHMENT HOSES to
the REPLENISHMENT TANKS.
• 2 REPLENISHMENT HOSES -
3
⁄
8
in. ID
• 2 REPLENISHMENT HOSE FITTINGS -
3
⁄
8
in. Barbed
• 2 REPLENISHMENT HOSE CLAMPS -
1
⁄
2
in.
(a) If the REPLENISHMENT TANKS are located in the MOUNTING STAND, connect the 2
REPLENISHMENT HOSES to the 2 REPLENISHMENT TANKS.
(b) If the REPLENISHMENT TANKS will be stored outside the MOUNTING STAND and outside the
darkroom, route the REPLENISHMENT HOSES from the PROCESSOR, under the sides of the
MOUNTING STAND, and to the REPLENISHMENT TANKS.
(c) If the REPLENISHMENT TANKS are located in the darkroom, do the steps below.
Rotate the MOUNTING STAND so that the open end of the STAND is positioned against the darkroom
wall.
Route the HOSES and POWER CABLE through the open end of the MOUNTING STAND and through
the holes in the WOOD FRAME.
To prevent light leaks, apply CAULK around the holes in the WOOD FRAME where the HOSES and
POWER CABLE pass through.
[50] Check that the STRAINER ASSEMBLIES are installed in the REPLENISHMENT HOSES near the
REPLENISHMENT TANKS.
[51] Install the POWER CABLE (provided with the M43A only) through the STRAIN RELIEF on the
PROCESSOR. If further detail is necessary, see “Making the Electrical Connection,” Steps a through f on
Page 50.
[52] Tighten the STRAIN RELIEF to secure the POWER CABLE.
Figure 20
Installing the POWER CABLE
120V, L-N
230V, L-N
240V, L-N
H130_0029HA
M43
(VIEWED FROM BELOW)
M43A/
CLINIC 1
STRAIN
RELIEF
WIRE 1
WIRE 2
TB1
L1
NEUTRAL
GROUND WIRE
E1
H130_0029HCA