INSTALLATION INSTRUCTIONS
52
September 1995 – 981088
[4] Check that the Water Usage Mode is correct. Refer to Table 1, “Setup Options” on page 37. The PROCESSOR
is factory set for Continuous Water Usage Mode. In this mode, water runs at one litre per minute. This allows
fresh water to circulate through the PROCESSOR and prevent biological growth. If biological growth is not a
concern, the PROCESSOR may be set to Low Water Usage Mode.
Note
If Low Water Usage Mode is configured, the WASH WATER DRAIN VALVE must be changed. A white dot on
the WASH WATER VALVE STEM indicates a valve with an oraface, which is used for Continuous Wash Mode
only. Refer to “Replacing the Drain Valves,” Plumbing Section in the Service Manual.
Note
If you are routing a water supply HOSE under the PROCESSOR, you will need a 90
°
fitting, not provided, at the
WATER INPUT SOLENOID connection.
Figure 44
Connecting the DRAIN HOSES and the HOSE from the Main Water Supply
H130_0026BA
fixer DRAIN HOSE
wash DRAIN HOSE
water supply HOSE
developer DRAIN HOSE
BIB FITTING
SOLENOID
WATER INPUT
H130_0026BCB