If LOCK AVAILABLE VALUES is set to NO, and the user selects SAVE SETTINGS from either the Voltage or
Current Stimulation Mode, STIM RATE 1 and PULSE DUR 1 will be overwritten with the saved values.
•
Since the VOLTAGE STIMULATION and CURRENT STIMULATION modes are available from the COOLED RF
AUTO TEMP READY and TRANSDISCAL TREATMENT READY modes, any changes made to the VOLTAGE
STIMULATION and CURRENT STIMULATION modes through ADVANCED SETTINGS also affects the
VOLTAGE STIMULATION and CURRENT STIMULATION modes when accessed through COOLED RF and
TRANSDISCAL modes.
Figure 7-20 ADVANCED SETTINGS Mode – STIMULATION SETTINGS Display
Settings:
Affected Mode
Parameter
DESCRIPTION
Available
Values
Default
Value*
LOCK AVAILABLE
VALUES?
Disable or enable selection of only
two STIM RATES and two PULSE
DURS. If “NO” is selected, the
Stimulation Preset Windows are
greyed out and not selectable.
Yes/No
Yes
STIM RATE 1
(Hz)
2 Hz
STIM RATE 2
(Hz)
Rate at which stimulus pulses are
delivered
1-Shot, 2, 5, 10,
20, 50, 75, 100,
150, 180, and
200 Hz
50 Hz
PULSE DUR 1
(ms)
1.0 ms
STIMULATION Modes
†
PULSE DUR 2
(ms)
Duration of pulses delivered
0.1, 0.2, 0.5, and
1.0 ms
0.1 ms
†
STIMULATION Modes refers to both VOLTAGE AND CURRENT STIMULATION Modes
7.7.2
ADVANCED SETTINGS Mode– LESION SETTINGS State
The following screen appears when LESION SETTINGS is selected from the ADVANCED SETTINGS Mode.
•
This screen allows adjustment of the initial RAMP TIME taken by the AUTO TEMP LESION and AUTO PULSED
LESION modes to attain the SET TEMPERATURE
•
This screen allows adjustment of POWER LIMIT in AUTO TEMP LESION and AUTO PULSED LESION modes
•
Changing RAMP TIME affects the portion of the total TIME spent at the SET TEMPERATURE
•
Lowering the RAMP TIME causes the Generator to attempt to reach the SET TEMPERATURE faster
•
MAX RAMP TIME EXTENSION is the maximum allowable duration for extension of the ramp time in Multi-RF
Auto Temp Lesion mode, to maintain a consistent Lesion Time. If all probes have not reached Set Temperature within
the sum of this time and the specified Ramp Time, an error notification will be generated. This feature can be turned
off by changing the setting to DISABLE.