025-0169-311 Revision: B
Page 36 of 52
Settings:
Affected Mode
Parameter
DESCRIPTION
Range/Units
Default
Value
Increment
PERIPHERAL
DISC WARNING
TEMP
Setpoint for the
Peripheral Disc
Temperature Warning
30–60 ºC
35 ºC
0.5 ºC
0.5 ºC/Min – 5 ºC/Min
0.1 ºC/Min
After 5 ºC/Min next
value is 10 ºC/Min
N/A
RAMP RATE
Rate at which
temperature increases
from the INITIAL
TEMP to the SET
TEMP
10–200 ºC/Min
2 ºC/Min
10 ºC/Min
Treatment Mode
POST
TREATMENT
COOLING
Select to enable or
disable the option to go
into POST
TREATMENT
COOLING after
treatment
ENABLE /DISABLE
ENABLE
Stimulation Mode
Placement Mode
Treatment Mode
NUMBER OF
PROBES
Number of TransDiscal
Probes used
1, 2
2
N/A
Treatment Mode
POWER LIMIT
Maximum power
available for
temperature control
1–50 W
15 W
1 W
7.7.5
ADVANCED SETTINGS Mode – RFA SETTINGS State
The following screen appears when RFA SETTINGS is selected from the ADVANCED SETTINGS Mode.
•
This screen allows the option of disabling the SECONDARY THERMOCOUPLE probe used in the RFA mode
o
If the SECONDARY THERMOCOUPLE is disabled, the corresponding temperature measurement window
in RFA – TREATMENT Mode will be inactive
•
This screen allows adjustment of the initial RAMP TIME taken by the RFA mode to attain the SET TEMP
Figure 7-24 ADVANCED SETTINGS Mode - RFA SETTINGS Display