025-0169-311 Revision: B
Page 41 of 52
8.5
Software Shutdown Limits During RF Delivery or Stimulation
•
Measured Impedance:
< 25
Ω
or > 3,000
Ω
•
Measured Temperature:
< 15
°
C, > 100
°
C, or
> SET TEMPE5
°
C for 5 s, or
> SET TEMPE10
°
C for 1 s
•
Measured RF Power:
> 55 W, or
> SET POWER + 50 % + 5 W for 100 ms, or
> SET POWER + 20 % + 5 W for 1 s
•
Measured RF Voltage:
> 180 V
rms
•
Measured RF Current:
> 0.9 A
rms
8.6
Hardware Shutdown Limits
•
Measured RF Power:
> 60 W
•
Measured RF Voltage:
> 200 V
rms
•
Measured RF Current:
> 1 A
rms
•
Stimulation Current:
> 40 mA
•
Stimulation Voltage:
> 11 V
8.7
Mechanical Specifications
•
Size:
9.5 x 12.5 x 14.0" (24 x 32 x 35 cm) maximum.
•
Weight:
16.5 lb (7.5 kg) maximum (not including power cord or shipping box)
•
Moisture protection rating:
IPX0 (ordinary, per IEC601)
8.8
Environmental Specifications
•
Operational temperature:
10 °C to 40 °C
•
Storage & Transportation temperature:
-40 °C to 70 °C
•
Operation, Storage & Transportation
Humidity:
20 % to 95 % non-condensing
8.9
Fuses
•
100 – 120 V~ / 60 Hz configuration:
Replace mains fuses as marked:
4A/250V, T-lag, 5x20mm.
•
220 – 240 V~ / 50 Hz configuration
Replace mains fuses as marked:
2A/250V, T-lag, 5x20mm
•
When replacing fuses, ensure the integrity of the new fuses by inspecting for physical damage that could affect the
function of the fuse
•
Use a precision slot drive screwdriver to remove the fuse drawer
•
Replace the fuses, and replace the fuse drawer
•
Ensure that the fuse drawer selection matches the value of the AC Mains Switch
o
WARNING: A mismatch of fuse type, fuse drawer selection and AC Mains Switch can result in permanent
damage to the Generator!
8.10
Line Input Ratings
•
3.5 A @ 100-120 VAC ~ 60 Hz
•
1.7 A @ 220-240 VAC ~ 50 Hz
8.11
Footswitch Specifications
•
Pneumatic