Note: Multi-RF Manual Power and Manual Pulsed modes support 1 probe connected to the Multi-RF module
Range
38–95 ºC
Increment
1 ºC
WARNING
TEMP
Default*
42 ºC
Range
10–900 s
Increment
1 s
TIME
Default*
120 s
Range
10–100 ms
Increment
10 ms
PULSE DUR
Default*
20 ms
Range
1–10 Hz
Increment
1 Hz
MANUAL
PULSED
LESION
ON
Temperature and
Impedance measurements
are displayed while Time
and Power are not displayed
(dashed lines replace
values).
In Multi-RF, the maximum
impedance is displayed
according to range:
R
rms
< 25
Ω
- ‘LOW’
25
Ω
< R
rms
< 3000
Ω
-‘OK’
R
rms
> 3000
Ω
- ‘HIGH’
VOLTAGE
STIMULATION
CURRENT
STIMULATION
AUTO TEMP
MANUAL
POWER
AUTO PULSED
MANUAL
PULSED
SAVE SETTINGS
While in the READY
state, other operating
modes for Baylis Pain
Management
Thermocouple probes
are available. When a
different mode is
selected, the Generator
will enter the READY
state of the new mode.
The present
SETTINGS can be
saved as defaults by
holding down the
SAVE SETTINGS key.
PULSE RATE
Default*
2 Hz