025-0169-311 Revision: B
Page 32 of 52
7.6
SAVE SETTINGS State
Figure 7-19 Save Settings Display
7.6.1
Example Display
•
When the Save Settings button is pressed and held for 5 seconds the setting values in the current mode will become
the new default values for those settings.
7.7
ADVANCED SETTINGS Mode
•
The ADVANCED SETTINGS mode is accessible only from the STANDBY state.
•
This state is entered by simultaneously pressing the upper-right Setting Adjustment Key [
] and the lower-left Mode
Select Key.
Displays and Indicators:
PARAMETER
STATUS
DESCRIPTION
RF/Stimulus Output
RF power is not delivered during the ADVANCED SETTINGS
mode.
FAULT Indicator
(The FAULT indicator is used only in the POST and SYSTEM
FAULT states.)
RF/Stimulus Output
ON/OFF Indicator
The indicator is not illuminated during the ADVANCED
SETTINGS mode.
Audio Output
OFF
Tone sounds briefly when saving settings.
Saving and Exiting:
SAVE SETTINGS *
Settings are stored, and the system transitions to the STANDBY mode.
CANCEL
The system transitions to the STANDBY mode without updating the settings.
*Defaults are factory-set, and are overwritten when the SAVE SETTINGS function is used.
7.7.1
ADVANCED SETTINGS Mode - STIMULATION SETTINGS State
When the ADVANCED SETTINGS mode is accessed, the Generator displays the default STIMULATION SETTINGS screen.
Settings on this screen apply to both VOLTAGE and CURRENT STIMULATION.
•
This mode allows the user to limit the quantity of selectable values in Stimulation Modes to two frequently used
values
•
STIM RATE and PULSE DUR may each be adjusted to toggle between two values that are frequently used during
treatment. The user may define these values here: STIM RATE 1 and STIM RATE 2, and PULSE DUR 1 and
PULSE DUR 2.
•
Toggling the STIM RATE in a Stimulation Mode will not toggle the PULSE DUR, and vice versa. The LOCK
AVAILABLE VALUES has only the function of limiting the list that is normally available for each setting to two
values.
•
If the LOCK AVAILABLE VALUES parameter is set to YES, it allows the user to toggle between the above two
stimulation rates and pulse durations. If it is set to NO, the user can view all values available for STIM RATE and
PULSE DURATION.
•
The values stored in STIM RATE 1 and PULSE DUR 1 are the default values upon entering Voltage or Current
Stimulation Mode.