SETTINGS Window: For each state (e.g. VOLTAGE STIMULATION, AUTO TEMP LESION,
TRANSDISCAL TREATMENT, etc.), the adjustable settings are shown in the SETTINGS Window. Up to 5
SETTINGS are adjustable in each state, by using the Setting Adjustment Keys (5) and Rotary Encoder (6), to the
right of the window.
(5)
Settings Adjustment Keys: The up
and down
buttons located to the right of the SETTINGS window (4)
are used for adjusting treatment settings in the upper 4 locations of the SETTINGS window.
(6)
Rotary Encoder: This encoder is used for adjusting the bottom parameter in the SETTINGS Window. The
parameter in the window depends upon the current mode.
(7)
OUTPUT ON/OFF Button and Indicator: This button controls STIMULATION, LESION, COOLED RF,
TRANSDISCAL, RFA and IDL outputs. In STIMULATION modes, this button toggles the output on and off.
In all other modes, this button control initiates or terminates RF output. The indicator is ON during RF delivery,
and also when in the STIMULATION: ON state.
(8)
POWER Indicator: This green LED indicates that system power is on.
(9)
FAULT Indicator: This red LED indicates a system fault has occurred. System faults include: self-test failure,
hardware protection (such as short-circuit shutdown), and software failure. Main power to the system must be
cycled (off-on) to attempt recovery from a system fault.
(10)
MEASUREMENTS Window: Measured values for elapsed TIME, probe TEMPERATURE(S), RF POWER,
RF VOLTAGE and/or IMPEDANCE are displayed during and after RF output in both STIMULATION and
LESION states and during treatment states for COOLED RF, TRANSDISCAL, RFA and IDL modes. During all
READY states (prior to RF output), TEMPERATURE and IMPEDANCE are displayed. In LESION READY
states for AUTOTEMPERATURE, AUTOPULSED, MANUAL POWER, AND MANUAL PULSED,
IMPEDANCE is displayed according to LOW/OK/HIGH range. In PLACEMENT states for TRANSDISCAL
and RFA modes, only the IMPEDANCE between the probe and the dispersive return electrode will be displayed.
(11)
Graph Window: This window is used for graphing temperature, power and voltage data acquired during RF
output in lesion states for ON states when in the LESION mode and treatment states when in COOLED RF,
TRANSDISCAL, RFA and IDL modes. The horizontal axis is time ( ), corresponding to the treatment duration
in the SETTINGS Window. The display will rescale if the TIME duration in the SETTINGS Window is
modified. The vertical axis represents temperature (
°
C) and either power (W) or voltage (V).
(12)
Connector Cable Connection: This patient-isolated connection is for attachment of the connector cable for
Baylis Pain Management Thermocouple probes.
(13)
Return Electrode Connection: This patient-isolated connection is for attachment of a Baylis Pain Management
Dispersive Electrode (PMA-GP-BAY).