025-0169-311 Revision: B
Page 37 of 52
Settings:
Affected Mode
Parameter
DESCRIPTION
Range/
Units
Default
Value
Increment
Treatment Mode
Placement Mode
SECONDARY
THERMOCOUPLE
Enable/disable optional
thermocouple
ENABLE/
DISABLE
ENABLE
N/A
Treatment Mode
RAMP TIME
Time in which
temperature increases
from the INITIAL TEMP
to the PEAK TEMP
15 – 60 s;
1 – 10 min
15 s
1 s (for 15-60 s);
30 s (for 1-10 min)
7.7.6
ADVANCED SETTINGS Mode - IDL SETTINGS State
The following screen appears when IDL SETTINGS is selected from the ADVANCED SETTINGS mode.
•
This screen allows the user to enable/disable the use of a secondary thermocouple probe as an aid by toggling
ENABLE/DISABLE on the SECONDARY THERMOCOUPLE setting
o
If the SECONDARY THERMOCOUPLE is disabled, the corresponding temperature measurement window
in IDL Mode will be inactive
•
This screen allows adjustment of temperature profile parameters for the IDL mode, which are not available from the
READY state. The INITIAL TIME, INITIAL TEMP, RAMP RATE, and POWER LIMIT are adjustable in this state
•
Changes to the INITIAL TIME and RAMP RATE affect the portion of the total TIME spent at the SET TEMP
•
Lowering the INITIAL TIME causes the Generator to attempt to reach the INITIAL TEMP more rapidly
Figure 7-25 ADVANCED SETTINGS Mode – IDL SETTINGS Display
Settings:
Affected
Mode
Parameter
DESCRIPTION
Range/
Units
Default
Value
Increment
SECONDARY
THERMOCOUPLE
Disable optional
Thermocouple Probe
ENABLE/
DISABLE
DISABLE
N/A
INITIAL TIME
Ramp-up interval to
INITIAL TEMP set point
15–120 s
30 s
1 s
INITIAL TEMP
Initial target for
temperature ramping.
40 ºC–80 ºC
65 ºC
1 ºC
RAMP RATE
Rate at which temperature
increases from the
INITIAL TEMP to the SET
TEMP
0.5 ºC/min –
5 ºC/min
2 ºC/min
0.1 ºC/min
IDL Mode
POWER LIMIT
Maximum power available
for temperature control
1–50 W
50 W
1 W
7.7.7
ADVANCED SETTINGS Mode – AUDIO SETTINGS State
The following screen appears when AUDIO SETTINGS is selected from the ADVANCED SETTINGS Mode
•
The AUDIO SETTINGS state in the ADVANCED SETTINGS mode allows the user to enable/disable the impedance-
related audio output during STIMULATION modes as well as during PLACEMENT and READY states (with
exception to IDL mode)