Figure 5-2 Generator rear panel
5.2
Rear Panel Displays, Controls, and Connections
•
Descriptions of rear panel features and their functions are given below. Please refer to Figure 5-2 for corresponding
locations on the rear panel.
(1)
Volume Adjustment: This knob adjusts the volume of the system’s audio output.
(2)
RS232 connection: This is an isolated connection to a standard 9-pin COM port. It is provided for passive data
acquisition, and cannot be used to control the system.
(3)
FOOTSWITCH connection: This pneumatic barb connects to the hose of the pneumatic footswitch. Like the
OUTPUT ON/OFF switch, the FOOTSWITCH controls stimulus output in the STIMULATION modes and
controls RF output in the LESION, COOLED RF, TRANSDISCAL, RFA and IDL modes. The action of the
FOOTSWITCH differs from the OUTPUT ON/OFF switch: the FOOTSWITCH must be held down to continue
delivery of stimulus pulses or RF energy.
(4)
Fuse Drawer: The orientation of the fuse drawer, in conjunction with the AC CONFIGURATION SWITCH,
determines the AC input voltage range (100-120V~ or 220-240V~). For each voltage setting the proper,
corresponding fuses must be used, as indicated in the rear panel labeling.
(5)
AC CONFIGURATION SWITCH: The orientation of this switch, in conjunction with the Fuse Drawer,
determines the AC input voltage range (100-120V~ or 220-240V~). Do not change the position of the voltage
selector while the system is plugged in.
(6)
Equipotential Ground Connection: This connector is attached to the chassis/earth ground. It is intended for
earth reference connection in environments where equipotential ground cabling is used.
(7)
AC MAINS SWITCH: This switch controls the initial AC power input to the system, and is part of the power
entry module, which integrates the AC power inlet and Fuse Drawer.
(8)
Fan: A brushless DC fan is used to exhaust warm air from the Generator. The direction of airflow is outward
from the rear panel.
(9)
Device Label: This label indicates the model number of the Generator and includes contact information for
Baylis Medical Company. The unique serial number of the Generator is marked on this label.
(10)
RWTUV Safety Mark Certification Label
(11)
CE 0120 CE Mark Label
(12)
RJ45 Connector: Connection to be used by authorized service personnel only.
(13)
Pump Module Interface Connector: For attachment of authorized cooling pump unit only.
(14)
WEEE Directive Label