025-0169-311 Revision: B
Page 29 of 52
PARAMETER
Measurement Window
Available Modes
Available Settings
STATE
Status
Description
Status
Description
Status
Description
Range
A, B, C or D
SELECT
PROBE*
Increment
N/A
Range
1-Shot–200 Hz
STIM RATE*
Available Values
†
1-Shot, 2, 5, 10, 20, 50, 75, 100, 150, 180, and 200 Hz
Range
0.1–1.0 ms
PULSE DUR
Available Values
†
0.1, 0.2, 0.5, and 1.0 ms
Range
0.0–10.0 V/mA
VOLTAGE
STIMULATION
CURRENT
STIMULATION
Impedance
measurements
are displayed.
VOLTAGE/
CURRENT
Increment
0.1 V/mA
Range
38–95 ºC
SET TEMP
Increment
1 ºC
Range
15–600 s
AUTO TEMP
LESION
TIME
Increment
1 s
Range
38–95 ºC
TEMP LIMIT
Increment
1 ºC
Range
15–600 s
TIME
Increment
1 s
Range
N/A
MANUAL
POWER
LESION
POWER LIMIT
Increment
N/A
Range
38–95 ºC
SET TEMP
Increment
1 ºC
Range
10–900 s
TIME
Increment
1 s
Range
10–100 ms
PULSE DUR
Increment
10 ms
Range
1–10 Hz
AUTO PULSED
LESION
PULSE RATE
Increment
1 Hz
Range
0–55 V
VOLTAGE
Increment
5 V (between 0 and 30V); 1 V (greater than 30V)
Range
38–95 ºC
WARNING
TEMP
Increment
1 ºC
Range
10–900 s
TIME
Increment
1 s
Range
20–100 ms
PULSE DUR
Increment
10 ms
Range
1–10 Hz
MANUAL
PULSED
LESION
ON
All
measurements
are displayed.
None
Other modes are
not selectable
while in the ON
state, as
indicated by the
greyed-out mode
key labels.
PULSE RATE
Increment
1 Hz