025-0169-311 Revision: B
Page 35 of 52
Settings:
Affected Mode
Parameter
DESCRIPTION
Range/Units
Default
Value
Increment
0.5 ºC/min – 5 ºC/min
0.1 ºC/Min
After 5 ºC/min, next
value is 10 ºC/min
N/A
RAMP RATE
Rate at which
temperature increases
from the INITIAL
TEMP to the SET
TEMP
10–200 ºC/Min
80 ºC/Min
10 ºC/Min
Cooled RF Auto
Temp Mode
POST TREATMENT
COOLING
Select to enable or
disable the option to
go into POST
TREATMENT
COOLING after
treatment
ENABLE/DISABLE
DISABLE
Stimulation Mode
Cooled RF Auto
Temp Mode
NUMBER OF
PROBES
Number of Cooled RF
Probes used
1, 2
1
N/A
Cooled RF Auto
Temp Mode
POWER LIMIT
Maximum power
available for
temperature control
1–50 W
12 W
1 W
7.7.4
ADVANCED SETTINGS Mode – TRANSDISCAL SETTINGS State
The following screen appears when TRANSDISCAL SETTINGS is selected from the ADVANCED SETTINGS Mode.
•
This screen allows adjustment of the warning temperature threshold for the Peripheral Disc Temperature (PDT) by
adjusting PERIPHERAL DISC WARNING TEMP
•
The Peripheral Disc Warning Temperature can be used as an aid during treatment. If either PDT A or PDT B exceeds
the Peripheral Disc Warning Temperature, the rate of the Output ON tone increases (the tone is played more often).
•
This screen allows adjustment of the RAMP RATE used by the TRANSDISCAL mode to attain the SET
TEMPERATURE
•
The option to use POST TREATMENT COOLING after the active treatment has ended can be accessed by toggling
the ENABLE/DISABLE Post Treatment Cooling control.
•
The option to use one probe is accessible by adjusting NUMBER OF PROBES.
•
The maximum amount of power available can be adjusted by rotating the dial to change SET POWER LIMIT.
Figure 7-23 ADVANCED SETTINGS Mode – TRANSDISCAL SETTINGS Display