025-0169-311 Revision: B
Page 18 of 52
7.3
READY States
•
VOLTAGE STIMULATION Mode – READY state is the default upon connection of a Baylis Pain Management
Thermocouple probe and cable.
Figure 7-5 VOLTAGE STIMULATION – READY Display
Figure 7-6 MULTI-RF VOLTAGE STIMULATION – READY Display
Displays and Indicators:
PARAMETER
STATUS
DESCRIPTION
RF/Stimulus Output
OFF
RF power is not delivered during the READY state. When Output ON/OFF is
pressed, RF power is delivered and the Generator automatically transitions to
the ON state.
VOLTAGE STIMULATION – During the ON state, Voltage must be greater
than 0.0 V for Stimulus pulses to be delivered.
CURRENT STIMULATION – Stimulus pulses are not delivered during the
READY state. During the ON state, Current must be greater than 0.0 mA for
Stimulus pulses to be delivered.
TRANSDISCAL AND COOLED RF TREATMENT - When Output ON/OFF
is pressed; the Generator will enter the PRE-TREATMENT COOLING state.
FAULT Indicator
OFF
(The FAULT indicator is used only in the POST and SYSTEM FAULT states.)
RF/Stimulus Output
ON/OFF Indicator
OFF
The indicator is not illuminated during the READY state, as there is no
Stimulus/RF power delivery.
RF Audio Output
None
RF audio output is not generated during the READY state.