025-0169-311 Revision: B
Page 30 of 52
PARAMETER
Measurement Window
Available Modes
Available Settings
STATE
Status
Description
Status
Description
Status
Description
Range
30–90 ºC
SET TEMP
Increment
1 ºC
Range
15s–30 min
COOLED RF
TREATMENT
TIME
Increment
15 s (between 15 s and 5 min)
30 s (between 5 min and 30 min)
Range
30–90 ºC
SET TEMP
Increment
1 ºC
Range
†
1–30 min
TRANSDISCAL
TREATMENT
All measurements are displayed.
TIME
Increment
30 s
Range
38–95 ºC
SET TEMP
Increment
1 ºC
Range
1–30 min
RFA MODE -
TREATMENT
All measurements are displayed. If the
Thermocouple Probe is disabled via RFA
SETTINGS state, accessible from the
ADVANCED SETTINGS mode, the
measurement is greyed out and the value
replaced with dashed lines.
TIME
Increment
30 s
Range
40–95 ºC
SET TEMP
Increment
1 ºC
Range
30 s–20 min
IDL MODE
ON
All measurements are displayed.
During ON state, SET TEMP should not be
decreased below INITIAL TEMP in
ADVANCED SETTINGS.
During ON state, SET TIME should not be
decreased below INITIAL TIME in
ADVANCED SETTINGS.
None
Other modes are
not selectable
while in the ON
state, as indicated
by the greyed-out
mode key labels.
TIME
Increment
30 s
* STIM RATE may not be changed until Voltage/Current set to 0 V/mA or until Output OFF; SELECT PROBE may not be changed until Output OFF.
†
Available values may be limited to two frequencies and two pulse durations if option selected in STIMULATION SETTINGS state, accessible from the
ADVANCED SETTINGS mode.