[SOURce:]CURRent Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
CURR:SST:RISE
Sets the soft start time when the output is turned on.
This command is valid when the startup state when the output is turned on is set to
CC priority (SYST:CONF:STAR:PRI CC).
Command
[SOURce:]CURRent:SSTart:RISE <numeric>
[SOURce:]CURRent:SSTart:RISE?
Parameter
Value:
0
Disable soft start (default)
0.5 to 99.9
Soft start time
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
S
Settings are reset to default when the *RST command is sent.
(Example)
CURR:SST:RISE 2500MS
Response
Returns the soft start time in NR3 format in response to CURR:SST:RISE?.
CURR:TRIG
Sets the current value that is applied when a trigger is sent.
When the PWR-01 is set so that constant current is controlled externally (CUR-
R:EXT:SOUR VOLT), settings performed by this command are invalid.
This command is invalid while a sequence is running.
Command
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <numeric>
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
Parameter
Value:
command
Unit:
A
Settings are reset to default when the *RST command is sent.
When you send the CURR command, this setting is set to the CURR value.
(Example)
CURR:TRIG 2.5
Response
Returns the current setting to be changed with a trigger in NR3 format in response
to CURR:TRIG?.