[SOURce:]CURRent Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
CURR:LIM:AUTO
Enables or disables the limit on the current setting.
If you enable the limit when the current setting is higher than 95 % of the OCP trip
point, the OCP trip point will be set to 105 % of the current setting.
This command is invalid while a sequence is running.
Command
[SOURce:]CURRent:LIMit:AUTO <boolean>
[SOURce:]CURRent:LIMit:AUTO?
Parameter
Value:
ON(1)
The current setting limit is enabled (default).
OFF(0) The current setting limit is disabled.
Settings are reset to default when the *RST command is sent.
(Example)
CURR:LIM:AUTO ON
Response
Returns whether the limit on the current setting is enabled in NR1 format in re-
sponse to CURR:LIM:AUTO?
Related Command
CURR:PROT
CURR:PROT:DEL
CURR:PROT
Sets the overcurrent protection (OCP) value.
This command is invalid while a sequence is running.
Command
[SOURce:]CURRent:PROTection[:LEVel] <numeric>
[SOURce:]CURRent:PROTection[:LEVel]?
Parameter
Value:
10 % to 112 % of the rated output current (Maximum by default)
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
A
Settings are reset to default when the *RST command is sent.
If you enable the current limit (CURR:LIM:AUTO ON) when the current setting is
higher than 95 % of the OCP trip point, the OCP trip point setting is changed to 105
% of the voltage setting.
(Example)
CURR:PROT 48
Response
Returns the OCP value in NR3 format in response to CURR:PROT?.
Related Command
OUTP:PROT:CLE
CURR:LIM:AUTO
CURR:PROT:DEL