[SOURce:]CURRent Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
PROG:UCOD
Sets the user code to identify sequences.
You can view the user code on the panel. To prevent unintentional sequence exe-
cution, we recommend that you set unique user codes for each sequence.
Command
PROGram[:SELected]:UCODe <NRf>
PROGram[:SELected]:UCODe?
Parameter
Value: 0 to 9999 (0 by default)
Settings are reset to default when the PROG:CLE command is sent.
(Example)
PROG:UCOD 7890
Response
Returns the user code in NR1 format in response to PROG:UCOD?.
[SOURce:]CURRent Command
CURR
Sets the current.
When the PWR-01 is set so that constant current is controlled externally (CUR-
R:EXT:SOUR CURR), settings performed by this command are invalid.
This command is invalid while a sequence is running.
Command
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <numeric>
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Parameter
Valut:
0 % to 105 % of the rated output current (Maximum by default)
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
A
Settings are reset to default when the *RST command is sent.
Rated output current
PWR401L
40 A
PWR801L
80 A
PWR1201L
120 A
PWR401ML
20 A
PWR801ML
40 A
PWR1201ML 60 A
PWR401MH
5 A
PWR801MH
10 A
PWR1201MH 15 A
PWR401H
1.85 A
PWR801H
3.70 A
PWR1201H
5.55 A
(Example)
CURR 2.5
Response
eturns the current setting in NR3 format in response to CURR?.