Appendix
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
Processing Time of Commands
The command processing time is the time until the next command is accepted.
The processing times indicated here are typical values. They are not guaranteed.
The processing times vary depending on the settings and the measurement condi-
tions.
It does not include the response time of the hardware.
Command
Processing Time (ms)
Description
USB RS232C
LAN
*1
Data rate
setting:
19200 bps
Data rate
setting:
1152000 bps
VXI-11 HiSLIP SCPI-
RAW
VOLT
CURR
0.72 4
0.7
2.4
0.7
0.02
Sets the
voltage or the
current
MEAS:VOLT?
MEAS:CURR?
0.72 15
2.8
4.75
2.4
2.8
Queries the
measured
value
OUTP
0.25 4
0.7
2.4
0.7
0.02
Output on/off
*IDN
0.34 25
4.5
4.75
2.2
2
Queries the
model name
*RST
190 3
0.4
2.4
0.7
0.02
Performs a
device reset
*1
: 100BASE-TX Ethernet
Errors related to the self-test function
These errors occur as results of self-tests executed with *TST? queries.
Error code
Error message description
+901 EEPROM MODEL info lost
Model information was lost.
+902 EEPROM CAL info lost
Calibration information was lost.
+903 Wrong Model ID setup
An incorrect model ID is specified.