Tutorial
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
For step 1, set the step time to 0.5 s and the current to 1 A.
PROG:STEP1:DWELL 0.5
PROG:STEP1:CURR 1
Set the user code to 1234. You can also view the user code on the panel.
PROG:UCOD 1234
Set the repetition count to infinity.
PROG:LOOP INF
After you have finished configuring the sequence, execute it.
■
Sequence example2 (triangular wave)
This section describes a sequence that continuously outputs a triangular wave.
In a step with voltage transition set to RAMP, the voltage changes linearly (actually,
step transition at 100 ms intervals) from the voltage setting of the previous step to
the specified voltage.
Parameter
Value
Repetition count Infinity
Interval loop
--
User code
5678
Parameter
Steps
0
1
Step time
1 s
1 s
Voltage
5.0 V
0.0 V
Voltage transition RAMP RAMP
Current
1.0 A
1.0 A
Current transition
IMM
IMM
Trigger output
Off
Off
Trigger input
Off
Off
First, reset.
*RST
Delete the current sequence, and create a new 2-step program.
PROG:CRE 2
You cannot change the number of steps later. If you want to add or delete them,
use the PROG:CRE command to set the number again.
The steps will contain default values (Step time: 1 s, Voltage: 0 V, Voltage transi-
tion: IMM, Current: Max., Current transition: IMM, Trigger output: Off, Trigger input:
Off).