120
PLZ-4WH
CC Mode Calibration (Calibration numbers 1, 2, and 3)
Perform calibration on the Low, Mid, and then High range. For each item, start calibration at
step A. Be sure to calibrate both offset and gain.
■
CC mode settings
The current settings for each model are listed below.
Calibrating the low range
■
Step A: Calibrate the offset values of the internal reference voltage for
current settings and of the measured values.
1
Connect a shunt resistor or ammeter that matches the value
corresponding to 10 % of the full scale low range value.
2
Press MENU (SHIFT+SET/VSET).
The menu screen appears.
3
Use
to select “3. Calibration,” and then press ENTER.
4
Use
and
to select “1. CC(Low).”
Current
Range
Percentage
of the Full
Scale (%)
Calibrated Items
Internal reference voltage
for current settings.
Measured Values
Internal Reference Voltage for
the Protection Function Settings
Offset
Gain
Offset
Gain
Offset
Gain
Low
10
Step A
–
Step A
–
Step B
–
100
–
Step C
–
Step C
–
Step D
Mid
10
Step A
–
Step A
–
Step B
–
100
–
Step C
–
Step C
–
Step D
High
10
Step A
–
Step A
–
Step B
–
100
–
Step C
–
Step C
–
Step D
Calibration Number and
Item
Output Setting of
the Power Supply
Current to Be Matched
PLZ164WH
PLZ334WH
PLZ1004WH
1
CC (Low)
Offset
Voltage:
3.3 V
Current:
Rated current of
the load device
8.25 mA±8.25 µA
16.5 mA±16.5 µA
50 mA±50 µA
Gain
82.5 mA±82.5 µA
165 mA±165 µA
0.5 A±0.5 mA
2
CC (Mid)
Offset
82.5 mA±82.5 µA
165 mA±165 µA
0.5 A±0.5 mA
Gain
825 mA±825 µA
1.65 A±1.65 mA
5 A±5 mA
3 CC (High)
Offset
825 mA±825 µA
1.65 A±1.65 mA
5 A±5 mA
Gain
8.25 A±8.25 mA
16.5 A±16.5 mA
50 A±50 mA
Memo
When calibrating multiple PLZ-4WHs of the same model operating in parallel operation, multiply the
current setting by the number of units.
Example: Three PLZ334WHs in parallel operation using CC (Mid) gain
(1.65 A±1.65 mA) x3=4.95 A±4.95 mA
When calibrating a setup that includes a PLZ2004WHB, multiply the rated power [kW]/1 [kW] to the current
setting for the PLZ1004WH.
Example: One PLZ1004WH (1 kW) and two PLZ2004WHBs (2kW) using CC (High) gain
Because the rated power is 5 kW, the current setting is (50 A±50 mA) x5=250 A±250 mA
Use shunt resistors and a regulated DC power supply whose rated currents are higher than the maximum
current that flows during calibration.