PLZ-4WH
23
1
In
st
al
la
ti
on and
Pr
ep
ar
at
ion
Connecting to the Load Input Terminals
The PLZ-4WH has load input terminals on both its front and rear panels. The specifications of
the product are for the load input terminals on the rear panel.
Connecting to the load input terminals on the rear panel
This section explains how to connect the DUT to the DC INPUT terminals on the rear panel
(the rear panel load input terminals).
Using the load input terminal cover
Pass the wiring that you intend to connect to the load through the load input terminal cover.
Cut the cover’s sleeves to match the thickness of the wiring
WARNING
Possible electric shock.
• Do not touch the load input terminals when the output is turned on.
• The load input terminals on the front panel are connected internally to the load input
terminals on the rear panel. The voltage applied to the terminal on one side is
transmitted directly to the terminal on the other side.
CAUTION
To avoid damaging the product, observe the following precautions.
• Do not connect the DUT to the DC INPUT terminals when the load is on.
• Do not connect electronic loads to the load input terminals on the front panel and those on
the rear panel at the same time.
To avoid overheating, observe the following precaution.
• Attach crimping terminals to the wires, and use the attached screw set to connect them.
WARNING
Possible electric shock. Always use the load input terminal cover.
10 mm diameter line
15 mm diameter line
20 mm diameter line
25 mm diameter line
Using the gauge as
a guide, cut the
sleeve appropriately
for the wire diameter.