128
PLZ-4WH
Specifications
Unless specified otherwise, the specifications are for the following settings and conditions.
•
The warm-up time is 30 minutes (with current flowing).
•
TYP:
These are typical values that are representative of situations where the
PLZ-4WH operates in an environment with an ambient temperature of 23
°
C.
These values do not guarantee the performance of the PLZ-4WH.
•
After PLZ-4WH has been warmed up, it must be calibrated correctly in a 23
°
C ± 5
°
C
environment according to the procedures given in the operation manual.
•
% of set: Denotes a percentage of the input voltage, input current, or input power setting.
•
% of f.s:
Denotes a percentage of the rated input voltage, rated input current, or rated input power.
•
% of rdng: Denotes a percentage of the input voltage, input current, or input power reading.
Ratings
Constant current (CC) mode
Model
PLZ164WH
PLZ334WH
PLZ1004WH
Operating voltage (DC)
1
1
The minimum operating voltage at which current begins to flow through the PLZ-4WH is approximately
50 mV. At the load input terminals on the rear panel.
5 V to 650 V
Current
8.25 A
16.5 A
50 A
Power
165 W
330 W
1000 W
Input resistance when the load is off
Approx. 2.21 M
Ω
2
2
Approx. 2.21/(the number of units) M
Ω
during parallel operation of units of the same model.
Model
PLZ164WH
PLZ334WH
PLZ1004WH
Operating
range
H range
0 A to 8.25 A
0 A to 16.5 A
0 A to 50 A
M range
0 A to 825 mA
0 A to 1.65 A
0 A to 5 A
L range
0 A to 82.5 mA
0 A to 165 mA
0 A to 500 mA
Setting range
H range
0 A to 8.6625 A
0 A to 17.325 A
0 A to 52.5 A
M range
0 A to 866.25 mA
0 A to 1.7325 A
0 A to 5.25 A
L range
0 A to 86.625 mA
0 A to 173.25 mA
0 A to 525 mA
Resolution
H range
300 µA
1 mA
2 mA
M range
30 µA
100 µA
200 µA
L range
3 µA
10 µA
20 µA
Accuracy of
setting
L range, M range
±(0.2 % of set + 0.1 % of f.s
1
)
1
The full scale of the range. However, for the M range, it is the full scale of the H range.
L range, 300
μ
A or more
±(0.2 % of set + 0.1 % of f.s)
L range, less than 300
μ
A
±(0.2 % of set + 0.1 % of f.s) + Vin
2
/2.21 M
Ω
2
Vin: The voltage at the load input terminals on the rear panel or sensing terminals
Parallel Operation
±(0.2 % of set + 1.1 % of f.s
)
Input voltage
variation
3
3
When the input voltage is changed from 5 V to 650 V at a current equal to the rated power/650 V.
H range, M range
20 mA
L range
2 mA
Ripple
rms
4
4
Measurement frequency bandwidth: 10 Hz to 1 MHz
2 mA
4 mA
12 mA
p-p
5
5
Measurement frequency bandwidth: 10 Hz to 20 MHz
20 mA
40 mA
120 mA