PLZ-4WH
95
Controlling Voltage Mode (CV mode and the +CV modes)
5
Ext
e
rn
al
Cont
rol
an
d Par
a
llel
Oper
at
ion
2
Connect the external voltage across pins 1 and 3 or pins 5 and 6 of the
EXT CONT connector.
If you are controlling the PLZ-4WH in the +CV modes, be sure to connect the external
voltage across pins 5 and 6.
3
Turn the POWER switch on, and make sure that the load is off.
4
Set the operation mode and the current range.
If you also want to control the current range externally, be sure to set the range to H.
5
If you have connected the external voltage across pins 1 and 3 of the
EXT CONT connector, from the menu, select “2.Configuration” >
“3.External” > “Control” > “V.”
If you have connected the external voltage across pins 5 and 6 of the
EXT CONT connector, from the menu, select “2.Configuration” >
“3.External” > “Control CV” > “ON.”
Doing so enables the external control of CV mode.
Setting Control to V and Control CV to ON from the menu enables the “Control CV”
that you have connected to pins 5 and 6 of the EXT CONT connector.
6
Turn the power switch off and then on again.
The menu settings are applied.
See