Constant Current Mode (CC mode)
2
Basi
c Funct
ions
Configuring CC+CV mode
CC+CV mode is constant current mode (CC mode) with CV mode added to it. You can add
CV mode even while the load is on.
1
Configure the constant current settings.
To configure the settings, refer to "Configuring CC mode".
2
Press +CV (SHIFT+MODE) to add CV mode.
The PLZ-4WH switches to CC+CV mode, and “CC+CV” appears on the display.
3
Make sure that the PLZ-4WH is in the fundamental setting entry
condition.
If SET is not highlighted, press SET/VSET to switch to the fundamental setting entry
condition.
4
Turn the rotary knob to set the current and voltage.
The value that you can set (current or voltage) changes when you press SET/VSET.
The CC+CV mode settings are complete. Press LOAD to start testing. You can change the
current and voltage even when the load is on.
SWITCHING
ENTER
SHORT
EDIT
SLEW RATE
RUN / STOP
UVP
PAUSE
STORE
RECALL
A
B
C
LOAD
Th/TL
FREQ / DUTY
MENU
PRESET SEQ
I
LEVEL
VRANGE
RANGE
SW ON
+CV
MODE
LOAD
SET / VSET
LOCK
SHIFT
ABC
OPP / OCP
Memo
When “SET” is
highlighted, the PLZ-4WH
is in the fundamental
setting entry condition.