116
PLZ-4WH
Inspection
To purchase accessories or options, contact your Kikusui agent or distributor.
Cleaning the dust filter
The dust filter is installed on the inside of the louver on the front panel. Periodically clean the
filter to prevent clogging.
1
Use both hands to grab the four edges
of the louver.
2
While pulling the bottom slat towards
you, slide the top of the louver down,
and remove the louver from the panel.
If the louver is difficult to remove, press down
on the top row to make the louver easier to
remove.
3
Remove the dust filter from the inside
of the louver and clean it.
Use a vacuum cleaner to dispose of the dust
and foreign particles that are attached to the
dust filter. If the filter is extremely dirty, clean it
using a water-diluted neutral detergent and
dry it completely.
4
Attach the dust filter to the louver.
Firmly attach the dust filter so that the louver’s
tabs protrude out of the top of the dust filter.
5
Attach the louver so that its tabs enter
the slots in the panel. While pushing
on the second slat from the bottom of
the louver, slide the louver up to attach
it to the PLZ-4WH.
CAUTION
• A clogged filter hinders the cooling of the inside of the PLZ-4WH, can cause malfunctions,
and can shorten the PLZ-4WH’s service life.
• When the PLZ-4WH is in operation, air is sucked through the dust filter to cool the inside of
the device.
If moisture is present in the dust filter, the temperature or humidity inside the PLZ-4WH
increases and may cause malfunctions.
Dust filter
Louver
Tab
±ᄻ
²ᄻ
ピン