E1412A 6 1/2 Digit Multimeter Service Guide
107
Execution Errors
Keysight E1412A Multimeter Error Messages
-160 to -168 Block data errors
The multimeter does not accept block data.
-170 to -178 Expression errors
The multimeter does not accept
mathematical expressions.
-211 Trigger ignored
A Group Execute Trigger (GET) or *TRG was received
but the trigger was ignored. Make sure the multimeter is in the “wait-for-trigger”
state before issuing a trigger, and make sure the correct trigger source is
selected.
-213 Init ignored
An INITiate command was received but could not be
executed because a measurement was already in progress. Send a device clear
to halt a measurement in progress and place the multimeter in the “idle” state.
-214 Trigger deadlock
A trigger deadlock occurs when the trigger source is
BUS and a READ? command is received.
-221 Settings conflict
This error can be generated in one of the following
situations:
–
You sent a CONFigure or MEASure command with autorange enabled and
with a fixed resolution. Example: CONF:VOLT:DC DEF,0.1
–
You turned math on (CALC:STAT ON) and then changed to a math
operation that was not valid with the present measurement function.
–
For example, dB measurements are not allowed with 2-wire ohms. The
math state is turned off as a result of this condition.
-222 Data out of range
A numeric parameter value is outside the valid range
for the command. Example: TRIG:COUN -3
-223 Too much data
A character string was received but could not be
executed because the string length was more than 12 characters. This error can
be generated by the CALibration:STRing and DISPlay:TEXT commands.
-224 Illegal parameter value
A discrete parameter was received which was
not a valid choice for the command. You may have used an invalid parameter
choice. Examples: CALC:FUNC SCALE (SCALE is not a valid choice) or
SAMP:COUN ON (ON is not a valid choice).
-230 Data stale
A FETCh? command was received but internal reading
memory was empty. The reading retrieved may be invalid or settings have
changed since the data was taken.
-330 Self-test failed
The multimeter’s complete self-test failed from the
remote interface (*TST? command). In addition to this error, more specific
self-test errors are also reported. See also “Self-Test Errors,” following this
section.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......