86
E1412A 6 1/2 Digit Multimeter Service Guide
Service
Repair/Maintenance Guidelines
Step 2
Place the module on its side and remove the torx screws that hold the
insertion/extraction handles and the PC assembly to the front panel. Separate the
front panel from the PC assembly.
Step 3
Place the module on its side with the address switch visible and remove
the eight T10 torx screws on the module’s shield.
Step 4
Lift the top shield off of the PC assembly to gain access to PC board
components. The PC board can be separated from the bottom shield.
Figure 8-2
Disassembly/Reassembly of Keysight E1412A