Keysight E1412A 6 1/2 Digit Multimeter
Service Guide
17
4
Verification Tests
Introduction
This chapter provides information on performance verification of the Keysight
E1412A multimeter. You can perform performance verification tests at three
different levels depending on your need:
–
Functional Verification Test - A series of internal verification tests
(self-tests) that give a high confidence that the multimeter is operational.
The self-tests take less than 20 seconds to complete.
–
Quick Verification Test - A combination of the internal self-tests and
selected verification tests (specified by Q in the verification tables). These
tests performed under computer control over the GPIB can be completed
in less than four minutes.
–
Performance Verification Test - An extensive set of tests that are
recommended as an acceptance test when the instrument is first received
or after performing calibration of the multimeter. These tests performed
under computer control over the GPIB can be completed in less than 20
minutes. Allowing time for thermal offset voltages to settle, on ranges
where thermal voltages impact on the measurement, can add another 10
minutes to the testing.
Recommended Test Equipment
Test equipment recommended for the performance verification and calibration
procedures are listed in Table 4-1. Use a source with accuracy requirements
indicated in the table for any substitute calibration standard.
300Vac (50 kHz) Testing
NOTE: The Fluke 5725A Amplifier is only necessary if you require testing the
300V, 50 kHz AC Voltage point (see Table 4-1).
Do not perform any of the following verification tests unless
you are a qualified, service-trained technician and have
read the WARNINGS and CAUTIONS in Chapter 1 (and the
Warnings and Safety information preceding chapter 1).
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...