E1412A Multimeter Specifications
To Calculate Total Measurement Error
Understanding the “% of range” Error
The
range error
compensates for inaccuracies that result from the function and
range you select. The range error contributes a constant error, expressed as a
percent of range,
independent
of the input signal level. The following table
shows the range error applied to the multimeter's 24-hour dc voltage
specification.
Total Measurement Error
To compute the total measurement error, add the reading error and range error.
You can then convert the total measurement error to a “percent of input” error or
a “ppm (part-per- million) of input” error as shown below.
% of input error
=
ppm of input error
=
Error Example
Assume that a 5Vdc signal is input to the multimeter on the 10 Vdc range.
Compute the total measurement error using the 90-day accuracy specifications:
(0.0020% of r 0.0005% of range).
Reading error= 0.0020% x 5Vdc
= 100mV
Range error= 0.0005% x 10Vdc
= 50mV
Total error= 100mV + 50mV
= ±150mV
= ± 0.0030% of 5Vdc
= ± 30 ppm of 5Vdc
Range
Input Level
Range Error
(% of range)
Range Error
Voltage
10Vdc
10Vdc
0.0004
40
V
10Vdc
1Vdc
0.0004
40
V
10Vdc
0.1Vdc
0.0004
40
V
Total Measurement Error
Input Signal Level
--------------------------------------------------------------
X 100
Total Measurement Error
Input Signal Level
--------------------------------------------------------------
X 1 000 000
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......