Keysight E1412A Digital Multimeter Service Guide
99
Frequency and Period Characteristics
E1412A Multimeter Specifications
Operating Characteristics
[5]
System Speeds [5]
Configuration Rates:
14/sec
Autorange Time:
<0.6 sec
Max. Internal Trigger Rate:
80/sec
Max. External Trigger Rate to Memory:
80/sec
[1] Specifications are for 1-hour warm-up at 100 PLC integration time.
[2] Relative to calibration standards.
[3] 20% overrange on all ranges, except 300Vac range which has 1% overrange.
[4] Input >100mV. For 10 mV input, multiply % of reading error x10.
[5] Speeds are for 0.02 PLC integration time, Delay 0 and 200Hz (Fast) ac filter.
Function
Integration Time
Readings/second
Frequency,
Period
100
1
1
9.8
0.02
80