INSTRUCTION MANUAL

DIGITAL MULTIMETER

KEW1021R

1. Safety Warnings

This instrument has been designed, manufactured and tested
according to IEC 61010: Safety requirements for electrical equipment,
and delivered in the best condition after passing quality control tests.
This instruction manual contains warnings and safety rules which have
to be observed by the user to ensure safe operation of the instrument
and to maintain it in safe condition. Therefore, read through these
operating instructions before using the instrument.

WARNING

● Read through and understand the instructions contained in this 

manual before using the instrument.

●  Keep  the  manual  at  hand  to  enable  quick  reference  whenever 

necessary.

● The instrument is to be used only in its intended applications. 
● Understand and follow all the safety instructions contained in the 

manual.

● It is essential that the above instructions are adhered to. Failure 

to follow the above instructions may impair the protection
provided by the instrument and test leads, and may cause injury,
instrument damage and/or damage to equipment under test.

The symbol

indicated on the instrument means that the user

must refer to the related parts in the manual for safe operation of the
instrument. It is essential to read the instructions wherever the
symbol

appears in the manual.

DANGER is reserved for conditions and actions that

are likely to cause serious or fatal injury.

WARNING is reserved for conditions and actions that

can cause serious or fatal injury.

CAUTION is reserved for conditions and actions that

can cause injury or instrument damage.

● Symbols listed below are used on this instrument. 

User must refer to the manual.

Instrument with double or reinforced insulation.

AC

DC

Ground (Earth)

This instrument complies to WEEE Directive (2002/96/EC).
Please contact your local distributor at disposal.

Measurement Category

O

Circuits which are not directly connected to the mains power supply.

CAT II

Primary electrical circuits of equipment connected to an AC

electrical outlet by a power cord.

CAT III

Primary electrical circuits of the equipment connected directly

to the distribution panel, and feeders from the distribution
panel to outlets.

CAT IV

The circuit from the service drop to the service entrance, and

to the power meter and primary over current protection
device(distribution panel).

This instrument is designed for CAT IV 300V/ CAT III 600V. Test
leads M-7066A with the supplied caps are designed for CAT
IV600V/ CAT III 1000V and without the caps are for CAT II 1000V.

DANGER

● Never make measurements under the circumstances exceeding the 

designed measurement category and the rated voltage of the
instrument.

● Do not attempt to make measurement in the presence of flammable 

gasses. Otherwise, the use of the instrument may cause sparking,
which can lead to an explosion.

● Never attempt to use the instrument if its surface or your hand is wet.
●  Do  not  exceed  the  maximum  allowable  input  of  any  measuring 

range.

● Never open the Battery compartment cover during a measurement. 
● To avoid electrical shock by touching the equipment under test or its 

surroundings, be sure to wear insulated protective gear.

●  Test  leads  to  be  used  for  voltage  measurements  shall  be  rated  as 

appropriate for Measurement Category III or IV according to IEC
61010-031 and shall have a voltage rating of 600V or higher.

● Barriers on the test leads provide protection to keep your fingers and 

hands from touching an object under test. Keep your fingers and
hands behind the barriers during measurement.

WARNING

● Never attempt to make measurement if any abnormal conditions, 

such as broken case and exposed metal parts are found on the
instrument or test leads.

● Verify proper operation on a known source before use or take action 

as a result of the indication of the instrument.

● 

Firmly attach the caps to the test leads when performing

measurements in CAT III or higher test environments. When
KEW1021R and the test leads are combined and used together,
whichever is lower category & voltage to earth either of them
belong to is applied.

● Do not rotate the Function Switch if the instrument and the   

equipment under test are connected.

● Do not install substitute parts or make any modification to the   

instrument. For repair or re-calibration, return the instrument to your
local KYORITSU distributor.

CAUTION

●  Use  of  this  instrument  is  limited  to  domestic,  commercial  and light 

industry applications. Strong electromagnetic interference or strong
magnetic fields, generated by large currents, may cause malfunction
of the instrument.

● Firmly insert the test leads. 
● Do not pull or twist the test leads to prevent the risk of damage. 
●  Power  off  the  instrument  after  use.  Remove  the  battery  if  the 

instrument is to be stored and will not be in use for a long period.

●  Do  not  expose  the  instrument  to  direct  sunlight,  high  temperature 

and humidity or dewfall.

●  Use  a  cloth  dipped  in  water  or  neutral  detergent  for  cleaning  the 

instrument. Do not use abrasives or solvents.

NOTE

● The LCD shows some digits at the Voltage or the Current range even 

while the test leads are open. And, it may show some digits instead

of 0 even if the test leads are shorted. However, these phenomena

don’t affect measurement results.

● A resistance measurement takes time to settle the reading if there is 

a high resistance or capacitance components.

2. Specification

● Accuracy (Temperature: 23 ± 5°C, Humidity: 45 - 75%) 

ACV

/

RMS

(Auto Range)

Range

Display Range

Accuracy (sine wave)

6V

0.000, 0.006 - 6.299V

±1.0 %rdg±3dgt

(40-500Hz)

60V

5.70 - 62.99V

600V

57.0 - 629.9V

Guaranteed accuracy : 0.01V-600V, less than 900V peak

Input impedance : approx. 10MΩ 

Hz Frequency -

ACV measurement

(Auto Range)

Range

Display Range

Accuracy

(sine wave)

99.99Hz

10.00 - 99.99Hz

±0.1%rdg±3dgt

999.9Hz

95.0 - 999.9Hz

9.999kHz

0.950 - 9.999kHz

99.99kHz

95.0 - 99.99kHz

Guaranteed accuracy : 10Hz-99kHz

% DUTY -

ACV measurement

Range

Display Range

Accuracy

(Square wave)

99.9 %

0.0 – 99.9 %

±1.0%rdg±3dgt

(50/60Hz)

Guaranteed accuracy : 10%-90%

DCV

(Auto Range)

Range

Display Range

Accuracy

6.000V

0.000 - ±6.299V

±0.5 %rdg±3dgt

60.00V

±5.70 - ±62.99V

600.0V

±57.0 - ±629.9V

Guaranteed accuracy : 0V-±600V

Input impedance : approx. 11MΩ(6V range) / 10MΩ(60/600V range) 

ACmV / RMS

Range

Display Range

Accuracy (sine wave)

600.0mV

0.0, 0.9 - 629.9mV

±2.0 %rdg±3dgt

(40-500Hz)

Guaranteed accuracy : 1.2mV-600mV, less than 900mV peak

Input impedance : approx. 900kΩ 

AC Clamp Sensor / RMS

(Auto Range)

Range

Display Range

Accuracy (sine wave)

60.00A

0.00, 0.09 - 62.99A

±2.0 %rdg±3dgt

+ Sensor accuracy (40-500Hz)

200.0A

57.0 - 209.9A

Direct reading from 10mV / A output Clamp sensor

Guaranteed accuracy : 0.12A-200A, less than 300A peak

Input impedance : approx. 900kΩ 

DCmV

Range

Display Range

Accuracy

600.0mV

0.0 - ±629.9mV

±1.5 %rdg±3dgt

Guaranteed accuracy : 0mV-±600mV,    Input impedance : approx. 900kΩ 

DC Clamp Sensor

(Auto Range)

Range

Display Range

Accuracy

60.00A

0.00 - ±62.99A

±1.5 %rdg±3dgt

+ Sensor accuracy

200.0A

±57.0 - ±209.9A

Direct reading from 10mV / A output Clamp sensor

Guaranteed accuracy : 0A-±200A,    Input impedance : approx. 900kΩ 

Resistance

(Auto Range)

Range

Display Range

Accuracy

600.0

Ω

0.0 - 629.9

Ω

±0.5 %rdg±4dgt

6.000kΩ 

0.570 - 6.299kΩ 

±0.5 %rdg±2dgt

60.00k

Ω

5.70 - 62.99k

Ω

600.0k

Ω

57.0 - 629.9k

Ω

6.000M

Ω

0.570 - 6.299M

Ω

40.00M

Ω

5.70 - 41.99M

Ω

±1.5 %rdg±3dgt

Guaranteed accuracy : 0Ω-40MΩ,    Open-loop Voltage : less than 3V 

Measurement current : less than 1mA

Continuity

Range

Display Range

Accuracy

600.0Ω 

0.0 - 629.9Ω 

Buzzer threshold value :

less than 90Ω 

Open-loop Voltage : less than 3V, Measurement current : less than 1mA

Diode

Range

Display Range

Accuracy

2.000V

0.000 - 2.099V

±5 %rdg±5dgt

Guaranteed accuracy : 0V-2V, Open-loop Voltage : less than 3V
Measurement current : approx.0.5mA (Vf=0.6V)

Capacitance

(Auto Range)

Range

Display Range

Accuracy

60.00nF

0.00 - 62.99nF

±2.0 %rdg±5dgt*

600.0nF

57.0 - 629.9nF

6.000µF

0.570 - 6.299µF

±5.0 %rdg±5dgt

60.00µF

5.70 - 62.99µF

600.0µF

57.0 - 629.9µF

1000µF

570 - 1049µF

* Accuracy after canceling floating capacitance using REL function.
Guaranteed accuracy : 0nF-1000µF

ACA / RMS

(Auto Range)

Range

Display Range

Accuracy (sine wave)

6.000A

0.000, 0.006 - 6.299A

±1.5 %rdg±3dgt

(40-500Hz)

10.00A

5.70 - 10.49A

Guaranteed accuracy : 0.01A - 10A, less than 15Apeak

Hz Frequency -

ACA measurement

(Auto Range)

Range

Display Range

Accuracy

99.99Hz

10.00 - 99.99Hz

±0.1 %rdg±3dgt

999.9Hz

95.0 - 999.9Hz

9.999kHz

0.950 - 9.999kHz

Guaranteed accuracy : 10Hz-9.9kHz

% DUTY -

ACA measurement

Range

Display Range

Accuracy

99.9 %

0.0 – 99.9 %

±1.0%rdg±3dgt (50/60Hz)

Guaranteed accuracy : 10%-90% (Square wave)

DCA

(Auto Range)

Range

Display Range

Accuracy (sine wave)

6.000A

0.000 - ±6.299A

±1.5 %rdg±3dgt

10.00A

±5.70 - ±10.49A

Guaranteed accuracy : 0A-±10A

●Measuring method : 

Σ modulation 

●Over-range indication : OL 
●Measurement cycle : 2.5 times per second 

(1000µF range of Capacitance function 0.05 times per second)

●Crest factor : less than 3 (45-65Hz) 

For non-sinusoidal waveforms, add ±0.5 %rdg±5dgt
(Applicable functions : ACV, ACmV, AC clamp sensor, ACA)

●Applicable standards : 

IEC 61010-1 /61010-2-033 : CAT IV 300V / CAT III 600V
Pollution degree 2, Indoor use, Altitude up to 2000m
IEC61010-31 (Test leads Model 7066A)
IEC 61326 (EMC) , EN 50581 (RoHS)

●Withstand voltage : AC5160Vrms 5sec between circuit and enclosure 
●IP rating : IP40 (IEC60529) 
●Insulation resistance : 100MΩ or more /1000V between enclosure   

and electrical circuit

●Operating temperature and humidity range : 

0 to 40°C, 80%RH or less (no condensation)

●Storage Temperature and humidity range : 

-20 to 60°C, 80%RH or less (no condensation)

●Power source : DC3V R03/LR03 (AAA) × 2 
●Current consumption : 3mA or less 
●Battery life : Approx. 200 hours (ACV, continuous, no load, with R03) 
●Dimension, Weight : 155(L)×75(W)×40(D)mm,   

approx. 250g (including batteries and Wing-type holder)

●Accessories : Test leads (M-7066A), soft case(M-9097) 

Instruction manual, 10A/600V Fuse (M-8919, included)
Battery R03 (AAA) 2pcs Flat-type holder, Wing-type holder

●Option : Magnet hanger strap (M-9189) 

Test leads with alligator clips (M-7234)
AC Clamp Sensor (KEW8161), AC/DC Clamp Sensor (KEW8115)

Function Switch

Cap

Barrier

REL / MAX

MIN Key

SELECT Key

Voltage Terminal

COM Terminal

HOLD / Back light Key

LCD

Current Terminal

3. Other Functions

● REL Function 

Press the REL key to enable this function and store the measured
value to display the differences between the stored value and the
values measured in further tests. The measurement range will be
fixed when the REL function is enabled, and the measuring range will
be between the initial value and the full scale value.

Press the REL Key again to release the stored value.

*To activate REL function, MAX/MIN function should be disabled.

● Max / Min value display function 

This function is to display the measured max and min values on the
LCD during a measurement. Press the MAX/MIN key 1 sec or longer
to start recoding of max and min values. Then the LCD shows the
latest max value. After that, the min and present values can be
toggled and checked by pressing the MAX/MIN key (less than 1
sec).

To disable this function, press the MAX/MIN key 1 sec or longer.

* To activate MAX/MIN function, REL function should be disabled.

● Data hold function 

Press the HOLD key (less than 1 sec). The LCD shows “H” mark and
the reading will be held (Data hold mode). The max/ min values are
not updated in MAX/MIN mode. Press the HOLD key
again (less than 1 sec) to release the display.

● Backlight function 

Press the Backlight key 1 sec or longer to turn
on the backlight. Press the Backlight key another 1 sec or longer to
turn it off. The light automatically turns off in 1 min.

● Low battery indication 

The LCD shows “

” mark when the batteries fall below the

normal operating voltage.

● Sleep Function

Automatically powers off the instrument in about 15 min after the last
switch operation.

To exit from the Sleep mode, rotate the Function switch or press any
key. To disable the Sleep function, press the HOLD/Backlight key

and power on the instrument.

Confirm that the LCD shows “P.OFF” about 1 sec.

●How to store test leads 

Attach the Wing-type holder to the back of the instrument to store the
test leads.

1. Attach the Wing-type holder to the back of the instrument.
2. Fit the barrier on the test lead into the groove between the

instrument and the Wing-type holder.

3. Then it will be easier to see the displayed readings during a

measurement.

● Other convenient functions 

4. ACV / DCV / ACmV / DCmV Measurement

DANGER

●  Before  starting  a  measurement,  always  check  and  confirm 

the Function switch is in the appropriate measurement
position and the test leads are connected to the proper input
terminals.

● Never make measurement on a circuit in which voltage over 

600V exists.

● Keep your fingers and hands behind the barrier during   

measurement.

4.1 ACV / DCV/ ACmV / DCmV measurement

(1) Set the Function switch to ACV, DCV, ACmV or DCmV position.

For frequency or DUTY measurement, set the switch to ACV
and press the SELECT key.

(2) Connect the test leads to the Voltage and COM terminals.

4.2 Clamp sensor (option) measurement

(1) Set the Function switch to ACmV or DCmV position and press the

SELECT key. LCD shows “SENSOR”.

(2) Connect the clamp sensor to the Voltage and COM terminals.

NOTE

● If the connection is reversed, the “ 

“ mark will be displayed on the

LCD. (DCV measurement).

● Press the REL key to adjust the reading on the DC clamp sensor to 

“0”.

● 

The sensor mode corresponds to direct reading with 10mV/A

output clamp sensor.

5. Resistance / Diode / Continuity / Capacitance

Measurement

WARNING

Never use the instrument on an energized circuit. Discharge the
capacitor before starting a capacitance measurement.

(1) Set the Function switch to the resistance position. For continuity

check or diode/ capacitance measurement, press the SELECT key.

(2) Connect the test leads to the Voltage and COM terminals.

NOTE

● LCD shows ”OL” when the test leads are open.   

(except for capacitance measurement)

● The LCD shows “OL” if the test lead connection is reversed for diode 

measurement.

● Measurement time on 600μF/1000μF range is a bit long.   

(20sec max.)

6. ACA / DCA Measurement

WARNING

The value for the maximum input current on ACA and DCA
ranges is 10A (protected by fuse). Do not apply current that
exceeds 10A.

If measuring a current of 6A or higher, the duration of

measurement should be within 2 min. After that, a rest
period of 10 min is required.

(1) Set the Function switch to ACA or DCA position. For frequency/

DUTY measurement, press the SELECT key while the Function
switch is in ACA position.

(2) Connect the test leads to the Current and COM terminals.

NOTE

● If the connection is reversed, the “ 

“ mark will be displayed on the

LCD. (DCA measurement).

● If the fuse is blown, please refer to 

7. Fuse Replacement

and replace

it with a new one.

7. Battery / Fuse Replacement

WARNING

● Replace the battery when the "            " mark- low battery voltage 

warning- appears on the LCD.
Otherwise, precise measurement cannot be made. If the battery
is completely exhausted, the LCD goes blank without showing
the "

" mark.

● Disconnect the test leads from the object under test and power 

off the instrument before opening the Battery compartment
cover for battery or fuse replacement.

●  The  fuse  gets  hot  after  current  measurement  and  may  cause 

burn injury. If replacing the batteries or fuse after current
measurement, a rest period of 10 min before you do.

● Use the designated fuse only. 
● Do not try to replace the battery or fuse if the surface of the   

instrument is wet.

CAUTION

● Do not mix old and new batteries.   
● Install a batteries in correct polarity as indicated in the   

Battery Compartment.

(1) Set the Function Switch to "OFF" position.
(2) Untighten the screw on the back of the instrument.
(3) Remove the Battery compartment cover and replace the

batteries or fuse.

(4) Attach the cover to the instrument and then secure the cover

by tightening the screw.

9-15

92-2238B

Magnet hanger strap

(option)

Press the SELECT key to toggle functions

Resistance

or Capacitance

Diode

Screw

Press the SELECT key to toggle functions

Black

Red

Battery

Fuse
(10A/600V, size : 6.3 × 32mm, Fast Acting type)

AC clamp sensor

DC clamp sensor

Black

Red

Press the SELECT key to toggle functions

Black

Red

Continuity

Default : ACV

Frequency

DUTY

Default:Resistance

Continuity

Diode

Capacitance

Default : ACA

Frequency

DUTY

Kyoritsu reserves the rights to change specifications or designs
described in this manual without notice and without obligations.

Black

Red

“Δ” appears and “AUTO” disappears 

when REL Key is pressed.

One press: REL ON
Another press: REL OFF

Max value

Min value

Present value

Less than 1 sec

Less than 1 sec

Less than 1 sec

1.

2.

Wing-type

holder

3.

Replace the batteries with

new ones when this mark

appears.

Flat-type

holder

Tilt stand

ACV / ACmV

DCV / DCmV

Default :ACmV

Sensor mode

(ACA)

Default :DCmV

Sensor mode

(DCA)

Press the SELECT key to toggle functions