background image

HOLDPEAK 

870CR

 

OPERATION MANUAL 

This LCD Auto Range & Auto Power off Digital AC/DC clamp 

multimeter is a portable, 3 3/4-digit multimeter. It is ideally suited 
for field, laboratory, shop and home applications. 

1. SAFETY INFORMATION

The following safety information must be observed to insure 
maximum personal safety during the operation at this meter. 
1) When measuring voltage ensure that instrument is not
switched to the current range, resistance range, diode and 
continuity range, capacitance range or temperature range. 
2) Use extreme care when measuring voltage above 50V.
especially from sources where high energy is existed. 
3) Avoid making connections to "live" circuits whenever possible.
4) Before making resistance measurements, diode or continuity
test, capacitance test or temperature test, ensure that the circuit 
under test isde-energized. 
5) Always ensure that the correct function and range is selected.
6) Extreme care should be taken when using the instrument to
conjunction with a current transformer connected to the terminals 
if an open circuit occurs. 
7) Ensure that the test leads and probes are in good condition
with no damage to the insulation. 
8) Take care not to exceed the over-load limits as given in the
specifications. 
9) Before opening the cover of the battery cabinet to replace
batteries. disconnect the test leads from any external circuit, set 
the selector switch to "OFF" position. 
10) Keep the fingers after the protection ring when measuring
through the instrument lead. 

1

1) Change the battery when the symbol “ ” appears to avoid

incorrect data. 

2. Panel Layout

1)

Rotary Switch: use this switch to select functions and ranges.

2)

D.HOLD/

 Back Light

 key

In any range, push the key, the

present display value will be locked and the " H " symbol will 
appear, push it again to exit HOL and the " H" symbol disappear.

 

Press “D.HOLD” button more than 2 seconds, the back light will light, 
press it more than 2 seconds again, the back light will light off.

 

3)

SELECT key: This key work on the "CAP

Ω" range only. 

Push the key to choose resistance, diode or continuity test.

 

4)

RANGE  key

Push  the  key  to  select  manual  mode,  push  it

again  to  change  the  range,  press  the  key  for  more  than  2 
seconds  to  go  back  auto  range  mode.  But  in  Hz/Duty  and 
Capacitance measurement, it can not select manual range mode.

 

5)

Hz/DUTY key

In Hz range, push the key, you can measure the

duty, push again, go back to    Hz    measurement. In voltage 
range, push it, you can measure Hz and duty, but the 
measurement range will be smaller, and    the auto range mode 
will be changed to manual range mode. 

6

REL key

Push the key, the present display value will be stored in 

memory, then the new display value is the difference between input 
value and stored data. In Hz/Duty measurement, it can not work. 

7)

LCD Display

LCD Dual Display, facilitates reads the data.

8)

T+

V Ω Input Jack 

COM    Input Jack

T-    Input Jack

2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS 

Display:    3 3/4 digit LCD with a max. reading of 3999. 
Range control:    Auto range control. 
Polarity:    Automatic negative polarity indication. 
Zero adjustment: Automatic. 
Over range indication: Only the "OL" display. 
Low battery: The "

" is display when the battery voltage is below 7.2V. 

Auto Power Off: 30 minutes after stopping the switch or no push 
button, the meter automatically enter to power off mode. Push 
button or run switch, auto power off disable. 
Safety Standards: The meter is up to the standards of IEC1010 

Double Insulation, Pollution Degree 2, Overvoltage Category 

Clamp opening size: 45mm. 
Operating Environment: Temperature 32

104

°

F(0

40

), 

humidity

80%RH. 

Storage Environment: Temperature -4

140

°

F(-20

60

),humidity

90%RH. 

Power supply: 9V Zinc-carbon battery. 

Dimension: 225(H)×77(W)×45(D)mm. 
Weight: Approx. 330g (including batteries). 

2.2 ELECTRICAL SPECIFICATIONS 

Accuracies are ±(% of rnumber in last digit) at 23

±

5

,

70%RH. 

2.2.1 DC Voltage 

Range 

Accuracy 

Resolution 

400mV 

0.8% of rdg+5digits 

0.1mV 

4V 

1.0% of rdg+5digits 

1mV 

40V 

10mV 

400V 

100mV 

1000V 

1.5% of rdg+5 digits 

1V 

Overload protection: 1000V DC/750Vrms AC 
Impedance: 10M

Ω,

More than 100M

Ω

  on 400mV scale 

2.2.2 AC Voltage 

Range 

Accuracy 

Resolution 

Frequency 

4V 

1.5% of rdg+5 digits 

1mV 

50

400Hz 

40V 

10mV 

400V 

100mV 

750V 

2.5% of rdg+5 digits 

1V 

50

100Hz 

Average sensing, calibrated to rms of sine wave 
Overload protection: 1000V DC/750Vrms AC;Impedance: 10M

Ω

2.2.3 DC Current 

Range 

Accuracy 

Resolution Frequency 

200A 

3.0% of rdg+10 digits 

0.1A 

1000A 

0

800 

3.5% of rdg+10 digits 

1A 

800

1000

 

6.5% of rdg+10 digits 

Overload protection: 1000Arms within 60 seconds 

2.2.4 AC Current

 

Range 

Accuracy 

Resolution Frequency 

400A 

3.0% of rdg+10 digits 

0.1A 

50

400Hz 

1000A 

0

800 

3.5% of rdg+10 digits 

1A 

800

1000

 

6.5% of rdg+10 digits 

Average sensing, calibrated to rms of sine wave 
Overload protection: 1000Arms within 60 seconds 

2.2.5

AC Voltage(True RMS) 

Range 

Resolution 

Accuracy 

Sensitivity 

400mV 

0.1mV 

±

2.0% of rdg+15 digits 

40Hz-1kHz 

4V 

1mV 

±

1.2% of rdg+10 digits 

40Hz-1kHz

 

40V 

10mV 

±

1.2% of rdg+10 digits 

40Hz-10kHz

 

400V 

100mV 

±

1.0% of rdg+10 digits 

40Hz-400Hz

 

750V 

1V 

±

1.0% of rdg+10 digits 

40Hz-400Hz

 

1. Input impedance : > 10M

Ω

  Ohm ACV True RMS.

2. Overload protect : 1000V DCV or ACV peak value

.

3. Press 

“Range” button for select ACV 400mV range.

2.2.6 Resistance 

Range 

Accuracy 

Resolution 

400Ω 

1.5% of rdg+15 digits 

0.1Ω 

4k

Ω 

1.2% of rdg+15 digits 

1Ω 

40k

Ω 

10Ω 

400k

Ω 

100Ω 

4

MΩ 

1k

Ω 

40MΩ 

2.5% of rdg+15 digits 

10k

Ω 

Overload protection: 250V DC/250Vrms AC 

2.2.7 Capacitance 

Range 

Accuracy 

Resolution 

40nF 

6.0% of rdg+10 digits 

10PF 

400nF 

3.5% of rdg+5 digits 

100PF 

4

μ

1nF 

40

μ

10nF 

100

μ

6.0% of rdg+10 digits 

100nF 

Overload protection: 250V DC/250Vrms AC 

2.2.8 Frequency 

Range 

Accuracy 

Resolution 

Sensitivity 

10Hz 

1.2% of rdg+10 digits 

0.01Hz 

Range of input 

voltage:1.5V~1

0V,

 

If input 

voltage over 

range, need 

adjust 

100Hz 

0.1Hz 

1000Hz 

0.8% of rdg+5 digits 

1Hz 

10kHz 

10Hz 

100kHz 

100Hz 

1000kHz 

1kHz 

10MHz  2.0% of rdg+10 digits 

10kHz 

Overload protection: 250V DC/250Vrms AC 