46
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Adjustments
Adjustment Procedures
Adjustment Procedures
Adjustment Conditions
See Table 4-1,
Recommended Test Equipment
, for test equipment requirements.
For optimum performance, all adjustment procedures should comply with
following test conditions:
–
Ambient temperature of the test area is between 18°C and 28°C and stable
to within ±1
o
C.
–
Ambient relative humidity of the test area is <80%.
–
Must have a two hour warm-up with all input signals removed.
–
Use only copper connections to minimize thermal offset voltages.
–
Shielded twisted PTFE insulated cable or other high impedance, low
dielectric absorption cable is recommended for all measurements.
–
Keep cables as short as possible. Long test leads can act as an antenna
causing the pick-up of ac signals and contributing to measurement error.
–
Allow 5 minutes after handling input connections for thermal offset voltage
settling.
–
For effective normal mode rejection (NMR), you must programmatically set
the proper power line frequency reference if you are not on a 60 Hz system
(e.g, 50 Hz or 400 Hz). The Keysight E1412A default is 60 Hz at power-on
and must be set to 50 Hz or 400 Hz if required using the CAL:LFR
command.
General Procedure
Before beginning any adjustment procedure, the multimeter MUST be in an
"UNSECURED" state. For assistance on how to accomplish this, see "Calibration
Security Code" earlier in this chapter. Each adjustment should be followed by a
performance verification test for added confidence. We recommend the
following general procedure always be followed.
–
UNSECURE the multimeter as outlined in the “Calibration Security Code”
section.
–
Perform the Zero Adjustment Procedure.
–
Perform the desired Gain Adjustment Procedure(s).
–
SECURE the multimeter and note the new calibration count in your
maintenance records. Also note any new secure code used.
–
Perform the Performance Verification Tests.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......