60
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Adjustments
Adjustment Procedures
–
A non-zero result indicates that an error has occurred. Send the
SYST:ERR? command and read the result. See Chapter 5 of the
Keysight
E1412A User’s Manual and SCPI Programmer’s Guide
for details.
Frequency Function Adjustment Point
Keysight E1412A Range (set with
FREQ:VOLT:AC:RANG)
Frequency Standard Output
1
1 Valid calibration value sent to the multimeter is any input with amplitude >100mV
rms and frequency of 1 kHz to 100 kHz.
1 V
1V @ 1
00
kHz
Figure 5-7
Frequency Adjustment Setup