Keysight E1412A 6 1/2 Digit Multimeter Service Guide
61
CALibration Command Summary
Adjustments
CALibration Command Summary
The CALibration command subsystem allows you to enter a security code to
prevent accidental or unauthorized calibrations of the multimeter. When
you first receive your multimeter, it is secured. You must unsecure it by entering
the correct security code before you can calibrate the multimeter (see
CALibration:SECure:STATe OFF|ON <code> command).
Subsystem Syntax
CALibration
:COUNt?
:LFRequency 5
0
| 60 | 400
:LFRequency? [MIN | MAX]
:SECure:CODE <
new code
>
:SECure:STATe OFF | ON
,
<code>
:SECure:STATe?
:STRing <quoted string>
:STRing?
:VALue <cal_value>
:VALue?
:ZERO:AUTO ON | OFF
:ZERO:AUTO?
:COUNt?
CALibration:COUNt?
queries the multimeter to determine the number of times a
point calibration has occurred. A complete calibration of the multimeter
increases the count by the number of points calibrated. It is not a record of
complete calibrations. The count is stored in non-volatile memory.
Comments
*RST does not change the calibration count stored in non-volatile memory.
Example
Querying the number of occurrences of point calibrations:
CAL:COUN
Query the calibration count.
:LFRequency
CALibration:LFRequency 50 | 60 | 400
sets the line frequency to either 50 Hz or 60 Hz.
Comments
–
The wrong line frequency setting will cause reading errors to occur.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......