E1412A 6 1/2 Digit Multimeter Service Guide
79
Troubleshooting
Service
610
DC gain x10 failed This test configures for the 1 V dc range. The dc
amplifier gain is set to x10. The measure customer (MC) input is selected to the
internal TSENSE source which produces 0.6 volts. A 20 ms ADC measurement is
performed and checked against a limit of 0.6 V ± 0.3 V.
611
DC gain x100 failed This test configures for the 100 mV dc range. The dc
amplifier gain is set to x100. The measure customer (
MC
) input is selected to the
internal
TSENSE
source which produces 0.6 volts. A 20 ms
ADC
measurement is
performed and checked for a +overload response.
612
Ohms 500 nA source failed This test configures to the 10 V dc range with
the internal 10 M 100:1 divider U102A connected across the input. The 500 nA
ohms current source is connected to produce a nominal 5 V signal. A 20 ms ADC
measurement is performed and the result is checked against a limit of 5 V ± 1 V
.
613
Ohms 5 uA source failed This test configures to the 10 V dc range with the
internal 10 M 100:1 divider U102A connected across the input. The 5 µA ohms
current source is connected. The compliance limit of the current source is
measured. A 20 ms ADC measurement is performed and the result is checked
against a limit of 7.5 V ± 3 V.
614
DC 1000 V zero failed This test configures to the 1000 V dc range with no
input applied. A 20 ms ADC measurement is performed and the result is checked
against a limit of 0 V ± 5 mV.
615
Ohms 10
A source failed This test configures to the 10 V dc range with
the internal 10 M 100:1 divider U102A connected across the input. The 10
A
ohms current source is connected. The compliance limit of the current source is
measured. A 20 ms ADC measurement is performed and the result is checked
against a limit of 7.5 V ± 3 V.
616
DC current sense failed This test configures to the 3 A dc range. A 20 ms
ADC measurement is performed and the result is checked against a limit of 0 A ±
5 A. This test confirms that the dc current sense path is functional. The test limit
is set wide because K101 does not open the current input during self-test. This
test should catch a dc current sense failure without causing false failures when
current inputs are applied during self-test.
617
Ohms 100
A source failed This test configures to the 10 V dc range with
the internal 10 M 100:1 divider U102A connected across the input. The 100
A
ohms current source is connected. The compliance limit of the current source is
measured. A 20 ms ADC measurement is performed and the result is checked
against a limit of 7.5 V ± 3 V.
618
DC high voltage attenuator failed This test configures to the 1000 V dc
range. The 500 nA ohms current source is connected to produce a nominal 5 V
signal. A 20 ms ADC measurement is performed and the result is checked
against a limit of 5 V ± 1 V.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......