40
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Verification Tests
Performance Test Record
Table 4-12
Keysight E1412A Multimeter Performance Test Record Date__________________
Test Input
DMM
Range
Low Limit
(Amps)
Measured Reading
High Limit
(Amps)
Meas
Uncert
Test Acc
Ratio
AC Current Gain Test
1A @ 1 kHz
1A
0.998100
1.001900
.00065
3:1
2A @ 1 kHz
3A
1.983200
2.016800
.0013
>10:1
Table 4-13
Keysight E1412A Multimeter Performance Test Record Date__________________
Test Input
DMM Range
Low Limit
(Hz)
Measured Reading
High Limit
(Hz)
Meas
Uncert
Test Acc
Ratio
Frequency Gain Test
10 mV @ 100 Hz
100 mV
99.90
100.10
.0005
>10:1
1V @ 100 kHz
1V
99990
100010
.5
>10:1
Measurement accuracy for resistance values applies to the Fluke
5700A displayed value NOT the nominal value. For example, a
programmed output of 1000
may actually output 1000.0231
.
Therefore, you must use the actual displayed output value for true
accuracy not the nominal value of 1000 or an additional error of
23 m
is introduced. The 24 hour specification for the 1k
range
is 25 m
and the additional 23 m
nearly compromises the
specification.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......